Warnings and Safety Precautions

– The TopMix meets the requirements for

electromagnetic compatibility (EMC).
Avoid exposing the equipment to
stronger interference than that specified
in the applicable standards (see “Decla-
ration of Conformity”).

– The casing on all connecting cables, 

as well as the casing on wires inside the
equipment housing, is made of PVC
materials. Chemicals that corrode these
materials must be kept away from these
cables.
All components of the TopMix (TM01-X)
paint-mixing weighing system are
restricted to an ambient operating tem-
perature range of 0°C to 40°C (32°F to
104°F). Make sure the place of installa-
tion is adequately ventilated to prevent
build-up of excessive heat.

!

Do not use electrical equipment out of
doors. Prevent build-up of static elec-
tricity; e.g., on the dust cover or glass
parts. Provide an adequate connection
from the TM01-X terminal and the Ex-
link box (YCO09-Z) to the equipotential
bonding conductor.

– Use the dust cover supplied to protect

the equipment from splashes of paint.
Observe the cleaning instructions when
cleaning the equipment.

– Have the equipment inspected at appro-

priate intervals for correct functioning
and safety by a trained technician.
Inspect the cables for damage regularly.

– Make sure operating personnel receive

sufficient training to recognize faulty
operating states and to introduce the
required measures in such cases (for
example, disconnecting the Ex-link box
from power).

Legal Notices

This equipment meets the prescribed
safety requirements. Improper use or
handling, however, can result in damage
and/or injury.
The manufacturer accepts no liability for
damage or injury resulting from failure
to observe these warnings and safety
precautions.

– When you use electrical equipment in

installations and under ambient condi-
tions subject to higher safety standards,
you must comply with the provisions as
specified in the applicable regulations
for installation in your country.

– The terminal and the Ex-link box

(YCO09-Z) are not permitted for use in
legal metrology, medical applications,
hazardous areas or areas in which poten-
tially explosive dusts are present. 

– The intrinsically safe terminal and the

Ex-link box (YCO09-Z) are manufactured
in compliance with the valid harmonized
European standards of CENELEC:
EN 50014:

1997 A1+ A2

EN 50020:

2004

The TopMix can be operated with 
intrinsically safe Sartorius weighing
instruments (e.g., PMA7500-X) in Zone 1
hazardous areas (see “Verification of
Intrinsic Safety,” 
Drawing No. 35960-740-60 A4 
or 

35960-740-61 A4)

– The industrial protection rating of the

terminal and the Ex-link box (YCO09-Z)
is IP40 in accordance with EN60529.
Handle the equipment carefully in keep-
ing with its IP rating. The place of use
must be secured accordingly.

Summary of Contents for TopMix TM01-X

Page 1: ...98648 011 50 Installation Instructions Sartorius TopMix Model TM01 X Electronic Paint mixing Weighing System...

Page 2: ...nnector for the power cord power supply 90 to 264 VAC 11 Connector for monitor VGA test monitor 800x600 pixels outside the hazardous area 12 Connector for the VGA video cable YCC01 0039M3 length 3 m C...

Page 3: ...can be connected to the Ex link box YCO09 Z outside the hazardous area Note Read the installation and operating instructions carefully before connecting the TopMix and putting it into operation 2 Gene...

Page 4: ...receive sufficient training to recognize faulty operating states and to introduce the required measures in such cases for example disconnecting the Ex link box from power Legal Notices This equipment...

Page 5: ...s equipment Use only cables and cable lengths approved by Sartorius The operator shall be solely responsible for any damage or injuries that occur when using cables not supplied by Sar torius The Ex l...

Page 6: ...inal TM01 X with LC display and keypad for use in Zone 1 hazardous areas Ex link box YCO09 Z for use outside the haz ardous area TopMix Ex link cable part no YCC01 0038M20 for connecting the terminal...

Page 7: ...g conductor is connected to the housing of the Ex link box All equipment must be connected to the equipotential bonding con ductor by connecting the grounding cable not included in delivery to the gro...

Page 8: ...Before connecting or disconnecting data cables or control lines to or from the device make sure all equipment is disconnected from power The explosion protected paint mixing weighing system must be s...

Page 9: ...length 20 m to the terminal Tighten the connector Connect the TopMix VGA video cable YCC01 0037M20 length 20 m to the terminal Tighten the fastening screws Connect the weighing instrument e g PMA scal...

Page 10: ...fied in national regulations for electrical installa tions Installation must be performed by a trained technician in accordance with national regula tions and acknowledged technological standards The...

Page 11: ...rews USB port see illustration Connect the TopMix PC USB cable YCC01 0040M3 length 3 m for a standard computer to the Ex link box outside the hazardous area PC side Connect the PC video cable YCC01 00...

Page 12: ...dous area PC side Computer setting high color 16 bit Connect the equipment to power PC side Note Observe the warnings and safety precautions included in these instructions Note The Ex link box YCO09 Z...

Page 13: ...ure the currently valid regulations and guidelines for installing equipment in a Zone 1 area in Germany ElexV of 27 Feb 1980 are strictly observed Switching the TopMix On and Off The TopMix terminal T...

Page 14: ...ervice contracts with regular maintenance intervals ranging from 1 month to 2 years The optimum mainte nance interval depends on operating conditions and tolerance requirements Repairs Defective equip...

Page 15: ...f you use electrical equipment in installa tions and under ambient conditions subject to higher safety standards you must com ply with the provisions as specified in the applicable regulations for ins...

Page 16: ...16...

Page 17: ...g 16 5 Ex link box approx kg 3 2 AC power source V 100 240V Frequency Hz 50 60 Power consumption average VA 25 Designation type of explosion protection Terminal TM01 X EEx ib d ib IIB T4 and II 2 G KE...

Page 18: ...Declaration of Conformity 18...

Page 19: ...19...

Page 20: ...20...

Page 21: ...21...

Page 22: ...22...

Page 23: ...s see note 5 Vmax Imax Ci Li appropriate for connection to associated apparatus with Entity Concept parameters listed above each circuit Vmax 12 6V Voc 12 0V Imax 33mA Isc 32mA Pi 103mW Po 94mW Ci 2 n...

Page 24: ...with Entity Concept parameters listed above 0 1 2 0 SARTORIUS AG 3 4 5 1 2 AC Supply 2 Video in AC DC 4 1 2 Video out 2 USB SARTORIUS AG 3 83 9 699 3 83 9 699 3 83 9 699 3 83 699 Cable 8 Type YCC01 00...

Page 25: ...ters 6 7 range 0 40 7 WARNING SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY 8 Maximum cable length 100m 328 ft 1 For installation Not for esters and alkanes atmospheres For 1 installation No...

Page 26: ...without the prior written permission of Sartorius AG The status of the information specifications and illustrations in this manual is indicated by the date given below Sartorius AG reserves the right...