17
Service Manual Signum
Working with the Service Mode
Additional Menus in the Service Mode
In the Text Menu (Setup)
In consecutive order following the menu items »
DATE
« and Code »
202122
«
– Service date »
S-DATE
«
– Memory number »
MEM-NO
«
– Service technician can enter a minimum load here »
SQMIN-S
«
– This displays or prints the minimum load »
SQMIN
«
In the number menu:
The setup menu is model-dependent for WP1 (
WP-1) and the interfaces COM1 (COM1)
and UniCOM (
UNICOM). It has been expanded to allow the following possible settings
for configuring weighing platforms WP-1(
WP-1) and WP2(WP-2):
1-9
Calibration/adjustment functions:
1-9-1: Ext. calibration/adjustment with a default weight (service mode not
required)
1-9-3: Ext. calibration/adjustment with a user-defined weight (service mode not
required)
1-9-4: Int. calibration/adjustment (service mode not required)
1-9-6:
1)
Internal linearization (only for WP-2 on COM1 and UniCOM)
1-9-7:
1)
External linearization with user-defined weights
1-9-8: Set preload (service mode not required)
1-9-9: Delete preload (service mode not required)
1-9-10: Key blocked (service mode not required)
1-9-18:
1)
Determine internal weight
1)
Only accessible in the service mode