56 Service Manual Signum
Working with the Service Software
Setting the XBPI-232 Protocol
You need to set the scale to the XBPI protocol (BPI-232) in order to perform
calibration/adjustment procedures or program the scale using the service
software.
– Switch the Signum scale off (press the
e key
).
– Switch the Signum scale back on (by pressing the
e key
). While all
segments are displayed, briefly press the
)
key.
APPL appears in the display.
– Using the
k
key, scroll to
SETUP.
– Press the
)
key.
WP-1 is displayed.
disp_wp-1_d.eps
disp_8888_d.eps
disp_appl_d.eps
disp_setup_d.eps