74
Calibration and Adjustment
User Manual Secura
®
, Quintix
®
, Practum
®
8.1 Calibration/Adjustment Using an Internal
Calibration Weight (Secura
®
and Quintix
®
only)
The balance is fitted with an internal calibration weight in the housing . This
calibration weight is automatically placed on the balance by a motor during internal
calibration and adjustment .
t
Ensure that the balance is stable, the draft shield is closed (if applicable) and the
weighing pan is empty .
t
Select
A
to zero the balance .
t
Select the menu key at the bottom left of the display in the current application .
Components
Statistics
Mixing
Weighing
y
The menu appears .
t
Select the CAL button to access the calibration function .
Calibrate balance
CAL-Extern
CAL-Intern
y
The
Calibrate balance
window appears .
t
Select
CAL-Intern
.
y
Calibration begins .
– The internal weight is automatically placed on the balance in the housing .
– The balance is calibrated and then immediately adjusted .
– The internal weight is automatically removed from the balance .
Date and Time:
14-Jun-2012 13:10
Calibration function:
Internal calibration
Start calibration:
Start: manually
Calibration deviation:
Dev 0.01 g
Adjustment function:
Internal adjustment
Adjustment deviation:
Dev 0.00 g
Report
y
A report appears on the display .
The report indicates the deviations identified during calibration . The report also
gives the result of the adjustment process .
t
Close the report window: Select
<
.
y
The balance is now calibrated and adjusted .
Information on potential errors can be found in the “Status Messages” chapter on
page 91 .
Summary of Contents for Practum
Page 1: ...98648 019 43 User Manual Secura Quintix Practum Laboratory Balances 98648 019 43...
Page 2: ......
Page 117: ...117 Specifications User Manual Secura Quintix Practum EC Type approval Certificate...
Page 118: ...118 Specifications User Manual Secura Quintix Practum Plates and Markings...
Page 119: ...119 Specifications User Manual Secura Quintix Practum Plates and Markings...