SCHEDULE
Canadian Certificate Of Conformity No: FM16CA0124X
THIS CERTIFICATE MAY ONLY BE REPRODUCED IN ITS ENTIRETY AND WITHOUT CHANGE
FM Approvals LLC. 1151 Boston-Providence Turnpike, Norwood, MA 02062 USA
T: +1 (1) 781 762 4300 F: +1 (1) 781 762 9375 E-mail:
[email protected] www.fmapprovals.com
F 348 (Mar 16)
Page 3 of 4
Ratings
- The PMA.Evolution and the YCO14 Ex-Link Converter are designed to be used together as a
system. The input voltage to the YCO14-Z Ex-Link Converter is via USB connections a 5 V ± 0.25 V. The
ambient operating temperature
range of the PMA.Evolution and YCO14 is 0°C to 40°C.
YCO14-Y Non-incendive Field Wiring Parameters:
Vmax = 17.0V, Imax = 2.0A, Ci = 5.7µF, Li = 0mH
Vmax = 16.0V, Ii Imax = 2.5A, Ci = 5.7µF, Li = 0mH
YCO14-Y “ic” Input Entity Parameters:
Ui = 17.0V, Ii = 2.0A, Ci = 5.7µF, Li = 0mH
Ui = 16.0V, Ii = 2.5A, Ci = 5.7µF, Li = 0mH
YCO14-Y Ex-Link Converter
YCO14-Z Ex-Link Converter
PMA.Evolution EVO1Xab-c
a = Up to three letters and/or numbers or blank (not critical to safety)
b = Up to two letters and/or numbers or blank (not critical to safety)
c = Up to five letters and/or numbers or blank (not critical to safety)
PMA.Evolution LAB1Xab-c
a = Up to three letters and/or numbers or blank (not critical to safety)
b = Up to two letters and/or numbers or blank (not critical to safety)
c = Up to five letters and/or numbers or blank (not critical to safety)
13.
Specific Conditions of Use:
YCO14-Y:
1. The YCO14-Y Ex-Link Converter shall only be used with the PMA.Evolution.
2. The protective ground connectors on the PMA.Evolution and the YCO14-Y Ex-Link Converter shall be
connected together using a potential equalisation conductor with a cross-sectional area of at least 4 mm
2
.
3. The CAT6 cable used between the PMA.Evolution and the YCO14-Y Ex-Link Converter shall be obtained
from Sartorius.
4. The enclosure of the YCO14-Y is non-conducting and may generate an ignition-capable level of
electrostatic charges under certain extreme conditions. The user shall ensure that the equipment is not
installed in a location where it may be subjected to external conditions that might cause a build-up of
electrostatic charges on non-conducting surfaces, additionally; cleaning of the equipment should be done
only with a damp cloth.
5. The equipment shall be installed in such a way that it is protected against UV radiation and the entry of
solid foreign objects or water capable of impairing the safety of the apparatus. Reduce the risk of
mechanical damage to a minimum.
6. If the Ex-Link Converter YCO14-Y is installed in Zone 2, the installation must be done in such a way that
the pollution degree 2 of the YCO14-Y board is guaranteed.
7. In Zone 2 the „USB Power“ port may not be used.
YCO14-Z:
1. The YCO-14Z Ex-Link Converter shall only be used with the PMA.Evolution.
2. The protective ground connectors on the PMA.Evolution and the YCO14-Z Ex-Link Converter shall be
EVO1X | LAB1X Installation Instructions
81