Data Input Format
You can connect a computer to your scale to send commands controlling weighing
instrument functions and applications via the interface port.
All commands use the same data input format. They begin with the ESC (ASCII:
27) character and end with a carriage return CR (ASCII: 13) and LF (ASCII: 10). The
total length of a command is anywhere from 4 characters (1 command character
between the start and end described above) to a max. of 7 characters (4 command
characters). This number can also be higher when sending texts.
The commands listed in the following table must each be supplemented with
ESC ... CR LF.
Example
The command character for output is “P“ (“output to Port“). To trigger this
command, send the string: “ESC P CR LF”.
Command Meaning
K
Weighing mode 1
L
Weighing mode 2
M
Weighing mode 3
N
Weighing mode 4
O
Block keys
P
Send display value to data interface
Q
Output acoustic signal
R
Unblock keys
D
Tare and zero
(combination tare function)
f3_
Zero (see also the “kZE_” command)
f4_
Tare without zeroing (see also the “kT_” command)
i_
Information about the indicator, example of output: “CAI/016302/1“
Meaning: Indicator: Combics 3, software version: 016302,
Active weighing platform: 1
kF1_
Trigger soft key F1 function
kF2_
Trigger soft key F2 function
kF3_
Trigger soft key F3 function
kF4_
Trigger soft key F4 function
kF5_
Trigger soft key F5 function
kF6_
Trigger
R
key function
kF7_
Trigger
H
key function
kF8_
Trigger
D
key function
kF9_
Trigger
k
key function
kF10_
Trigger
J
key function
kF11_
Trigger
K
key function
kF12_
Trigger
L
key function
kP_
Trigger
p
key function (print at printer interface)
kT_
Trigger
T
key (tare)
kNW_
Trigger
n
key function (toggle the weighing platform)
kZE_
Trigger
(
key function (zero the instrument)
kCF_
Trigger
c
key function
x1_
Output model designation of active weighing platform, example: “LP6200S-0C“
x2_
Output serial number of active weighing platform, example: “0012345678“
x3_
Output software version of active weighing platform, example: “00-43-04“
x4_
Output software version of indicator, example: “01-63-02“
x9_
Output serial number of indicator, example: “0012345678“
x10_
Output model of indicator, example: “CAIS3“
z1_
Input: printout header 1
z2_
Input: printout header 2
The ASCII code for the “underline” character (“_”) is 95.
Format for entering printout header lines: “ESC z x a ... a _ CR LF“ with x=1 or 2
and a ... a: 1 to 20 characters for header x, followed by the underline, CR and LF
characters.
txx…x_
Write text in display. xx…x is the text to be displayed.
Operating Instructions Combics Complete Scales
101
Data Interfaces
Summary of Contents for Combics CAH3
Page 130: ...130 Operating Instructions Combics Complete Scales Declarations of Conformity ...
Page 131: ...Operating Instructions Combics Complete Scales 131 Declarations of Conformity ...
Page 132: ...132 Operating Instructions Combics Complete Scales Test Certificate ...
Page 133: ...Operating Instructions Combics Complete Scales 133 Plates and Markings ...
Page 134: ...134 Operating Instructions Combics Complete Scales Plates and Markings ...
Page 135: ...Operating Instructions Combics Complete Scales 135 Plates and Markings ...
Page 136: ...136 Operating Instructions Combics Complete Scales Plates and Markings ...
Page 139: ...Operating Instructions Combics Complete Scales 139 ...
Page 144: ......
Page 146: ......
Page 147: ......