![Sartorius combics 1 Operating Instructions Manual Download Page 37](http://html1.mh-extra.com/html/sartorius/combics-1/combics-1_operating-instructions-manual_1199617037.webp)
37
Example:
Initialize the Classification application by taking over weighed values (menu item 4.9.1) to define 3 classes (menu item 4.81.)
c
Delete any existing initialization data
O
Begin initialization
The minus sign of the bar graph flashes to prompt the weight for the upper limit
of the first class. A “1" is shown in the numeric display.
Place the weight for the upper limit of the first class on the weighing instrument
(in this example, 10.0 g)
O
Store the weight value as the upper limit for the first class
The middle segments of the bar graph flash to prompt the placement
of the weight for the upper limit of the second class on the weighing instrument.
A “2" is shown in the numeric display.
Place the weight for the upper limit of the second class on the weighing instrument
(in this example, 20.0 g)
Combics 2
OK
REF
CF
Fn
0
T
n
-
Combics 2
OK
REF
CF
Fn
0
T
n
-
Summary of Contents for combics 1
Page 108: ...Dimensions Scale Drawings All dimensions are given in millimeters 108 49 90 187 302 ...
Page 113: ...113 ...
Page 114: ...114 ...
Page 115: ...115 ...
Page 116: ...116 ...
Page 117: ...117 ...