Both interfaces cannot be used simultaneously because the RS232 port will be disconnected 
automatically after plugging the USB cable into the port. 
 
 

Alert switch 

 
Both, Radon Scout and Radon Scout PLUS are equipped with a switch output which can be 
used for ventilation control or alert indication. The potential-free contacts of the switch will be 
connected at the end of the sampling interval if the measured Radon concentration exceeds 
the  threshold.  The  output  will  be  deactivated  one  minute  before  the  subsequent  interval 
ends.  Therefore,  the  sampling  interval  must  be  set  at  least  to  two  minutes.  The  alert 
threshold is fixed at 300Bq/m³ for the Radon Scout while  it is user adjustable  in the Radon 
Scout PLUS. The connector is located on the rear panel of the instrument. In case of pending 
alert, the red light at the front panel will blink instead of the green one. 
 
As  an  accessory,  we  offer  a  wireless  power  switch  which  can  be  used for  direct  ventilation 
control. The  switch  includes  a  transmitter  unit  to  be  connected  to  the  alert  switch  output of 
the  instrument,  and  a  plug  adaptor  (receiver)  which  can  be  placed  between  the  wall  outlet 
and the power cable of the ventilation unit. Optionally, a flush-mounting power switch is also 
available.  The  plug  adaptor  offers  an  additional  timer  function  to  define  the  duration  of  the 
ventilation  period  independently  on  the  sampling  interval  of  the  instrument  (for  example  15 
minutes  on  while  the  sampling  interval  is  60  minutes).  The  wireless  switch  is  coded  -  that 
means,  several  transmission  lines  can  be  established  within  one  building.  The  assignment 
between transmitter and receiver is easily done by a push button.  
The sampling interval should be set to 60 or 120 minutes if the threshold level is lower than 
300Bq/m³. 
 

Technical Data 

 
Alert Switch 

 

 

Max. switchable voltage 

24V 

 

Max. switchable current 

0,4A 

 

Max. pulse current 

1,5A 

 

connector 

2-pins 

Manufacturer: Binder, series  710 

Wireless switch 

 

 

Transmitter 

2 inputs 

Additional input can be used for manual 
switch 

Frequency 

433MHz 

With coding for secure transmission 

Range 

approx. 30m 

 

Power supply 

Battery CR2032 

Good for 20000 switch actions 

Plug adapter 

220V/50Hz 

6 code memories for group switching 
(e.g. for more than one Radon monitor) 

Max. switchable power 

3500W 

 

Timer settings 

7s, 5min, 30min, 
1h, 2h, 4h, 8h 

If the timer function is not used the 
power switch follows the alert switch 

  
 
 
 

Summary of Contents for RADON SCOUT

Page 1: ...port sarad de Internet www sarad de CONTENT THE INSTRUMENT 2 GENERAL 2 FRONT PANEL ELEMENTS 2 POWER SUPPLY 3 START OF A MEASUREMENT 4 INTERRUPTION OF A MEASUREMENT 4 TO PERFORM A MEASUREMENT 4 Prepara...

Page 2: ...an external power supply are required Therefore use and exposition at home or at workplace is possible without any disturbance A total duration of measurement of up to four month with continuos data r...

Page 3: ...ng message appears on the display of the RADON SCOUT PLUS Stored data remain in the memory and can be read out after changing of discharged batteries The selection of the right battery depends on the...

Page 4: ...nish the first integration interval until the first value can be calculated is displayed As soon as the results of the first integration interval are available the display switches over automatically...

Page 5: ...y mode stopped It will be possible to start a measurement and stop it one time After setting the button into STOP position the sampling will be finished and the result of the whole test period appears...

Page 6: ...nstrument and a plug adaptor receiver which can be placed between the wall outlet and the power cable of the ventilation unit Optionally a flush mounting power switch is also available The plug adapto...

Page 7: ...r 0 100 rH Pressure Sensor 800 1200 mbar Alert switch 24V 0 4A potential free Opto MOS relais Tamper sensor 2G to detect movements or shocks Memory 672 records RADON SCOUT PLUS 16383 Power supply 2 x...

Reviews: