SARAD RADON SCOUT User Manual Download Page 4

SARAD

 

RADON-SCOUT 

 

Manual_Radon-Scout_EN_24-03-17.doc 

If the cell voltage drops further below 0.95 Volts, the measurement is stopped automatically 
and the device enters into the stand-by modus. Discharged batteries are indicated by the red 
LED, flashing each four seconds. 
 
 

Start of a measurement 

 
Setting  the  toggle  switch  to  the  right  starts  the  measurement  of  the  Radon  activity 
concentration  in  the  ambient  air.  Older  data  will  remain  in  the  memory  and  can  transferred 
later together with the new one. 
 

RADON-SCOUT only 

In case that the switch is already set to the position “RUN” during replacement of discharged 
batteries,  the  measurement  will  be  started  immediately.  Please  set  the  real  time  clock  and 
restart  the  measurement  once  again.  The  LED  will  flash  in  green  for  about  half  a  second, 
then LED will flash one or three times each four seconds dependent on selected integration 
interval (one flash = three hours, flashing three times = 1 hour). 
 

RADON-SCOUT-PLUS only 

 
If  the  toggle  switch  is  already  in  „RUN“  position  after  replacing  the  battery,  the  display  will 
show a request to set the real time clock of the instrument by the PC software. 
To start the acquisition, the s

witch has to be set firstly to “STOP” and then to “RUN” position. 

The green LED lights for a short time and blinks each 4 Seconds during acquisition. 
After starting, the remaining time to finish the first integration interval (until the first value can 
be calculated) is displayed. 
As soon as the results of the first integration interval are available, the display switches over 
automatically  between  two  display  pages  (each  Minute).  The  first  page  contains  the  actual 
value  for  Radon  with  statistical  1-Sigma  error  as  well  as  the  sensor  readings  (T,  rH,  p), 
battery voltage and time stamp. The second page shows the Radon average since the last 
start  of  an  acquisition,  the  time  of  the  start,  the  time  period  from  start  (used  for  averaging) 
and the number acquired records. 
All results are stated in the physical units as selected by software (SI or US style). 
  
 

Interruption of a measurement  

 
A  recent  measurement  series  may  be  interrupted  by  switching  the  toggle  switch  to  the 
position “STOP”, assumed that it is not locked by the software. The Radon-Scout will enter 
the “STOP” modus immediately without finishing the recent integration interval. 
The RADON-

SCOUT PLUS will show “Radon-Scout PLUS” 

 
 

To perform a measurement 

 

Preparation 

 
Prior  to  each  new  measurement  campaign,  the  status  and  type  of  the  batteries  should  be 
checked to avoid a loss of data with respect to a interruption of the analysis. In case that the 
cells had been changed after you adjusted the real-time clock the last time 

– adjust the timer 

again to ensure a reliable time regime!  
 

Selection of the right location for exposure 

 
The  Radon-Scout  was  designed  for  stationary  monitoring.  Strong  vibration,  movement  or 
mechanical  shocks  can  disturb  the  analysis.  This  interference  will  be  detected  and 

Summary of Contents for RADON SCOUT

Page 1: ...port sarad de Internet www sarad de CONTENT THE INSTRUMENT 2 GENERAL 2 FRONT PANEL ELEMENTS 2 POWER SUPPLY 3 START OF A MEASUREMENT 4 INTERRUPTION OF A MEASUREMENT 4 TO PERFORM A MEASUREMENT 4 Prepara...

Page 2: ...an external power supply are required Therefore use and exposition at home or at workplace is possible without any disturbance A total duration of measurement of up to four month with continuos data r...

Page 3: ...ng message appears on the display of the RADON SCOUT PLUS Stored data remain in the memory and can be read out after changing of discharged batteries The selection of the right battery depends on the...

Page 4: ...nish the first integration interval until the first value can be calculated is displayed As soon as the results of the first integration interval are available the display switches over automatically...

Page 5: ...y mode stopped It will be possible to start a measurement and stop it one time After setting the button into STOP position the sampling will be finished and the result of the whole test period appears...

Page 6: ...nstrument and a plug adaptor receiver which can be placed between the wall outlet and the power cable of the ventilation unit Optionally a flush mounting power switch is also available The plug adapto...

Page 7: ...r 0 100 rH Pressure Sensor 800 1200 mbar Alert switch 24V 0 4A potential free Opto MOS relais Tamper sensor 2G to detect movements or shocks Memory 672 records RADON SCOUT PLUS 16383 Power supply 2 x...

Reviews: