INSTALLATION
Installation site
To operate this unit properly and to obtain maximum performance, install the unit in a location with the
following conditions:
1. A location not subjected to direct sunlight
Installation in a location subjected to direct sunlight may lead to inadequate cooling.
2. A location with adequate ventilation
Leave at least 10 cm around the unit for ventilation. Poor ventilation will result in a reduction of the
refrigeration capacity.
3. A location away from heat generating sources
Avoid installing the unit near heat-emitting appliances such as gas ranges or stoves. Heat can cause
inefficient refrigeration.
4. A location with a sturdy and level floor
WARNING
Install the unit on a sturdy floor.
If the floor is not strong enough or the installation site is not
adequate, this may result in injury from the unit falling or tipping over.
Select a level and sturdy floor for installation.
This precaution will prevent the unit from tipping.
Improper installation may result in water spillage or injury from the unit tipping over.
5. A location where objects will not fall on the unit
CAUTION
Never install the unit in a place where the possibility exists that an object may fall on it.
There are
cooling circuits and electric components in the upper part of this unit. Current leakage or electric shock
may be caused if an object falls on it.
6. A location without flammable or corrosive gas
WARNING
Never install the unit in a flammable or volatile location.
This may cause explosion or fire.
Never install the unit where acid or corrosive gases are present
as current leakage or electric shock
may result due to corrosion.
7. A location not prone to high humidity
WARNING
Do not use the unit outdoors.
Current leakage or electric shock may result if the unit is exposed to
rain water.
Never install the unit in a humid place or a place where it is likely to be splashed by water.
Deterioration of the insulation may result which could cause current leakage or electric shock.
CAUTION
The unit starts defrosting frequently
due to excessive frost on the evaporator if it is installed in high
temperature and high humidity location.
The chamber temperature goes up to approximately 10
o
C
temporarily during defrosting.