OPERATING INSTRUCTIONS
^ - ( Supplementary note )-
For120V, 280Vunits :
Lowering the sterilization temperature setting by 2°C(3.6°F)or more after the chamber
has been heated causes the temperature indication to flash on and off.
For 230V units :
Lowering the sterilization temperature setting by 4°C(7.2°F) or more after the chamber
has been heated causes the temperature indication to flash on and off.
This is not a malfunction, and the flashing will stop as soon as the temperature in the
chamber drops to the new sterilization temperature setting.
•Sterilization time setting
Follow the steps below to change the initial sterilization
time to a new time setting. The allowable setting range
is 1-180 minutes.
(1)Press the timer setting Q ^ i ] b u t t o n . The current steri-
lization time setting appears on the digital indicator
lamp.
(2)Press the
button to change the time
setting with holding down theDM13button. (The steri-
lization time setting can be raised or lowered in 1-minute
increments between 1 and 30 minutes and in 5-second
increments between 30 and 180 minutes. Holding down
the ffllia or JMWWiSI button for more than one second
causes the setting to change faster.)
The setting of the sterilization temperature and steriliza-
tion time are remained in memory as long as the unit is
used continuously. If the power is turned off, the settings
are lost, however, if the unit is powered on again, the
settings return to the initial settings.
Supplementary note j-
The initial temperature and time settings (memory pat-
tern) can be sellected as follows by specifying an internal
setting in the main unit.
Pattern
1
2
3
4
Sterilization temperature
1 2 1 °C(250°F)
1 2 1 °C(250°F)
1 1 5°C(239°F)
1 1 0°C(230°F)
Sterilization time
15 minutes
20 minutes
15 minutes
15 minutes
The above four options are available. Contact your deal-
er to select a pattern.
TEMP./TIMER
I HIGH PRESSURE
DON'T TURN THE HANDLE, IF
THIS DISPLAY IS FLASHING.
LIGHTING UP : STERILIZING
FLASHING COOLING
TEMP
TIMER
A UP
VDOWN