D-1034-141 SANWA PC520M(E)
— 38 —
RANGE
Accuracy
Burden Voltage
50Hz - 60Hz
500.0µA
0.15mV/µA
5000µA
0.6% rdg +3dgt
0.15mV/µA
50.00mA
3.3mV/mA
500.0mA
1.0% rdg +3dgt
3.3mV/mA
5.000A
0.6% rdg +3dgt
0.03V/A
10.00A*
0.03V/A
40Hz - 1kHz
500.0µA
0.15mV/µA
5000µA
0.8% rdg +4dgt
0.15mV/µA
50.00mA
3.3mV/mA
500.0mA
3.3mV/mA
5.000A
1.0% rdg +4dgt
0.03V/A
10.00A*
0.03V/A
AC Current
*10A continuous
Ω
Ohms
RANGE
Accuracy
50.00
Ω
0.4% rdg+ 6dgt
500.0
Ω
0.2% rdg+ 3dgt
5.000k
Ω
, 50.00k
Ω
, 500.0k
Ω
0.2% rdg+ 2dgt
5.000M
Ω
1.0% rdg+ 3dgt
50.00M
Ω
1.5% rdg+ 5dgt
Open Circuit Voltage : <1.3VDC (<3VDC for 50
Ω
& 500
Ω
ranges)
Temp Temperature ( & )
RANGE
Accuracy*
-50 TO 1000
0.3% rdg + 3dgt
*K type thermocouple range & accuracy not included
PC520M̲英 06.4.28 3:31 PM ページ 38