Sanosil Q-Jet Superior
36
6.2
Regular visual check and Maintenance
Do not tip the device on its side at any time unless completely empty and dry.
Group/Component
Operation
Cycle
Air circuit /
Device structure
Airfilter:
replace every 10-20hours
monthly
Structure:
visual check of the state
(dents, breakages of parts)
monthly
Device in general
Visual control:
Checking operationality
daily
Components:
status of device’s components
(integrity of components, fluid loss, sealings,
caps)
monthly
Documentation:
Verification that all documentation
for the device is updated and available
yearly
Hydraulic Circuit
Nozzle:
Visual status check (cracks, points of rupture,
brui- sing, occlusions, deposits)
daily
Hydraulic Circuit
Peristaltic pump tubing: check / replace
two-yearly
6.3
Extraordinary Maintenance
Intervention-Type
Inspection
Checking / Verification
Visual Inspection
Leakage and Seals
Flow and Internal Tensions
A
X
X
X
B
X
X
X
C
X
X
X
D
X
X
6.3.1
Inspection / Check after an intervention or assistance involving replacements or repairs of components.
6.3.2
Inspection / Check after device has not been used for a longer period (6-12 months) or after it has been on
work for 2 years.
6.3.3
Inspection / Check after change of person responsible for the device.
6.3.4
Inspection / check to be made yearly.