Sanosil Q-Jet Superior
23
5.11
Program Normal with timer settings
Starts the whole room disinfection process to a programmed day and time
Please read and follow important safety directions on page 6 before use!
Settings
„
program normal with timer
“
•
Starting on main screen, press
„F2“
or icon
„Settings“
to input
users‘
settings
•
Select desired program type (Normal)
Oder
•
Note for users using a printer (type of printer page 42).
Please note: After every operation with the Sanosil Q-Jet
a printout can be printed by pressing the “reprint” button.
•
Touch fields:
Volume/Concentration/Code Charge/Typology to
open the respective keyboard screen.
Input or change values according to your needs and confirm your
inputs with Enter.
- Volume:
room volume in m
3
(length x width x height)
-
Concentration:
dosage desired per m
3
(usually 8, 12, 16ml)
-
Code Charge:
Definition of disinfection liquid
- Typology:
Name of room (e.g. surgery room 2)
•
For Timer: press
“Down
Arrow“
(next page)
•
Press button of desired week day.
F2