Electromagnetic immunity
Guidance and manufacturer’s declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS
The Blood pressure monitor (SBC 53) is intended for use in the electromagnetic environment specified below. The customer of the
user of the Blood pressure monitor (SBC 53) should assure that it is used in such and environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile. If floor are
covered with synthetic material, the relative humidity should be
at least 30%.
Power frequency
(50Hz) magnetic field
IEC 61000-4-8
3A/m
3A/m
Power frequency magnetic fields should be at levels characte-
ristic of a typical location in a typical commercial or hospital
environment.
Summary of Contents for SBC 53
Page 72: ...72 1 Sanitas SBC 53 1 72 2 73 3 76 4 77 5 79 6 82 7 83 8 87 9 87 10 88 11 90...
Page 74: ...74 5 30...
Page 75: ...75 30 3...
Page 76: ...76 WEEE Waste Electrical and Electronic Equipment 3 1 2 3 4 5 M1 M2 M1 M2 3 4 2 1 M1 M2 5...
Page 78: ...78 Pb Cd Hg Bluetooth Bluetooth M1 M2 5 M1 M2 M1 M2 M1 M2...
Page 80: ...80 1 5 3...
Page 81: ...81 9 M1 M2 3 Bluetooth Bluetooth 30 Bluetooth 30 Bluetooth 30 HealthCoach HealthCoach 7...
Page 82: ...82 3 Bluetooth 3 5 6 5...
Page 84: ...84 M2 Bluetooth M1 M2 1 M1 2 M2 A M1 M2 AM 7 5 00 9 00...
Page 85: ...85 M1 M2 PM 7 18 00 20 00 M1 M2 03 M1 M2 30 A 5 M1 M2 Bluetooth...