Appendix II
The Sphygmomanometer (MD4200) is intended for use in the electromagnetic environment specified below. The
customer or the user of the Sphygmomanometer(MD4200) should assure that it is used in such an
environment.
Electrostatic discharge (ESD)
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Electrical fast transient/burst
IEC 61000-4-4
Power supply lines: ±2 kV
Power supply lines: ±2 kV
Surge
IEC 61000-4-5
line(s) to line(s): ±1 kV 100 kHz
repetition frequency
line(s) to line(s): ±1 kV 100 kHz
repetition frequency
Power frequency magnetic field IEC
61000-4-8
N/A: UT is the a.c. mians voltage prior to application of the test level.
Remark *. The EUT do not contain magnetically sensitive components or circuitry. So this test do not need to evaluated.
30 A/m
50Hz/60Hz
30 A/m
50Hz/60Hz
Proximity Magnetic Fields
IEC 61000-4-39
30KHz, CW,8A/m
134.2KHz - PM 2.1KHz, 65A/m
13.56MHz - PM 50KHz, 7.5A/m
N/A
Conduced RF
IEC 61000-4-6
150KHz to 80MHz:
3Vrms
6Vrms (in ISM and amateur radio bands) 80%
Am at 1kHz
150KHz to 80MHz:
3Vrms
6Vrms (in ISM and amateur radio bands) 80%
Am at 1kHz
Voltage dips, short interruptions and
voltage variations on power supply
input lines
IEC 61000-4-11
0% 0.5 cycle
At 0˚, 45˚, 90˚, 135˚, 180˚, 225˚, 270˚ and 315˚ 0
% 1 cycle And 70% 25/30 cycles Single phase:
at 0 0% 300 cycle
0% 0.5 cycle
At 0˚, 45˚, 90˚, 135˚, 180˚, 225˚, 270˚ and 315˚ 0
% 1 cycle And 70% 25/30 cycles Single phase:
at 0 0% 300 cycle
Immunity test
IEC 60601-1-2 Test level
Compliance level
Guidance and manufacturer’s declaration - electromagnetic Immunity
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz - 2,7 GHz
80% AM at 1 kHz
10 V/m
80 MHz - 2,7 GHz
80% AM at 1 kHz