S&A CW-5300 User Manual Download Page 11

TY-JS-20319-003

11

(2) Exhaust Air

After adding water for the first time or replacing water, exhaust the air in the water pump
to start use, otherwise the equipment will be damaged. The exhaust method is as follows

Method 1: Under the state of shutting down, after adding water, remove the water outlet

and connect the water pipe, drain for 2 minutes, and then install it firmly.
Method 2

Open the water supply inlet, after starting the machine (water flows),

repeatedly press and fold the water pipe several times to drain the air from the pipe.

8.Test-run Check.

(1) Check whether the pipeline connection is correct, and there must be no bubbling or

water leakage;

(2) Check whether the inlet and outlet water ball valves are open and the drain valve is

closed;

(3) Check the liquid level of the water tank;

(4) Confirm that the electrical wiring of the equipment is connected correctly;

(5) Confirm whether the equipment is grounded.

V. Operating States &Parameters Adjustment

The new T-506D intelligent temperature controller is designed with intelligent control

mode. It will self-adjust controlling parameters according to room temperature for meeting

equipment cooling requirements.

The new T-506E intelligent temperature controller is designed with constant temperature
control mode with water temperature maintained at 25

.Users can adjust it as needed.

T-506D and T-506E controllers are of same functions and structure except factory setting.

1.Temperature control panel introduction

2.Indicators of temperature controller working status

Refrigeration indicator - During refrigeration process, this indicator is on. In constant

temperature control mode, this indicator is off. During the delay process, this indicator
flashes.

Refrigerant solenoid valve indicator - When connected, this indicator is on. When

disconnected, this indicator is off.

Heating indicator - When connected, this indicator is on. When disconnected, this

indicator is off.

Alarm indicator - When connected, this indicator is on. When disconnected, this

indicator is off.

3.Quick setting

Press “SET” key to set the temperature. The upper window indicates F0 (F1 is indicated
in intelligent control mode) while the lower window flashes with current value. Then press

or

key to change the value which will be memorized by the temperature

Summary of Contents for CW-5300

Page 1: ...TY JS 20319 003 1 CW 5300 6000 6100 6200 6260 Industrial Chiller User Manual...

Page 2: ...rly so that you can refer to it whenever you need information This manual is not a quality guarantee Our company reserves the right to the interpretation of the correction of misprint and improperly d...

Page 3: ...4 I Overview 7 II Model Illustration 7 III Product Outlines Parts 8 IV Installation 9 V Operating States Parameters Adjustment 11 VI Alarm Output Terminal 14 VII Maintenance 15 VIII Simple Troublesho...

Page 4: ...on a The power supply must conform to the standard indicated in the nameplate or the manual b Use the standard cable and select the standard wire diameter c The grounding wire must be installed and th...

Page 5: ...ilable Otherwise it may cause anesthesia and suffocation if a large amount of refrigerant fills the enclosed space b Avoid contacting with skin or it will cause frostbite DANGER Failure to take safety...

Page 6: ...r filter at least once a week If it is blocked the cooling ability will decrease and the power consumption will increase and the alarm will not work properly Equipment surface cleaning a Please use cl...

Page 7: ...hiller the refrigerant in the evaporator coil absorbs the heat of the return water and vaporizes into steam The compressor continuously extracts the generated steam from the evaporator and compresses...

Page 8: ...mes are only for one of the models The outlines and parts installation may be slightly different in different models Power Switch Temperature Controller Air Inlet Filter Gauze Water Supply Inlet Elect...

Page 9: ...etc 4 Operating Environment Ambient Temperature Ambient Humidity Altitude 0 40 90 RH 3 000m 5 Medium Requirements Cooling medium allowed by the chiller includes purified water distilled water high pu...

Page 10: ...1 Please select the cable based on the max rated current indicated on the label of the chiller 2 Recommended reference standard for power cord diameter selection Rated current A 5 10 15 25 35 50 Wire...

Page 11: ...control mode It will self adjust controlling parameters according to room temperature for meeting equipment cooling requirements The new T 506E intelligent temperature controller is designed with con...

Page 12: ...ings After the power is on for 30 seconds press and hold and keys for 3 seconds until it displays rE The supplier parameter and user parameter will restore to factory value 3 seconds later it will ret...

Page 13: ...perature is 5 higher than the set water temperature and the alarm of over low water temperature will be on when water temperature is 10 lower than the set water temperature Users can easily adjust wat...

Page 14: ...rly disassemble the filter gauze for cleaning and use a compressed air gun to blow away the dust on the condenser about 1 week is suggested and the air filter must not be missing for a long time Displ...

Page 15: ...ending deformation Check water circulation pipe Ultrahigh water temperature alarm Blocked dust gauze bad thermolysis Disassemble and clean the dust gauze regularly Poor ventilation for air outlet and...

Page 16: ...TY JS 20319 003 16 Attachment Schematic CW 5300 flow sensor CW 5300 flow switch...

Page 17: ...TY JS 20319 003 17 CW 6000 flow sensor CW 6000 flow switch...

Page 18: ...TY JS 20319 003 18 CW 6100 6200 flow sensor CW 6100 6200 flow switch...

Page 19: ...TY JS 20319 003 19 CW 6260 flow sensor...

Reviews: