background image

EXCEEDING MEASUREMENT RANGE IN BODY MODE:

•  When measurement value is lower than 95°F (35°C), it displays Lo (Figure 4.1) and gives 4 consecutive beeps and the backlight on the LCD 

screen will turn red.

•  When measurement value is higher than 109.2°F (42.9°C), it displays Hi (Figure 4.2) and gives 4 consecutive beeps and the backlight on the 

LCD screen will turn red.

EXCEEDING MEASUREMENT RANGE IN OBJECT MODE:

•  When measurement value is lower than 32.4°F (0.2°C), it displays Lo (Figure 4.3) and gives 4 consecutive beeps and the backlight on the LCD 

screen will turn red.

•  When measurement value is higher than 199.4°F (93°C), it displays Hi (Figure 4.4) and gives 4 consecutive beeps and the backlight on the 

LCD screen will turn red.

NOTICE:

•  When surrounding temperature is lower than 60.8°F (16°C) or higher than 104°F (40°C), it displays Err (Figure 4.5) and gives 4 consecutive beeps 

and the backlight on the LCD screen will turn red. Under this condition, it is not advised to use this thermometer and accuracy is not assured.

PRODUCT NUMBER: THDG986

Non-Contact Infrared Human 

Forehead Thermometer Instructions

MEASUREMENT SETTINGS CONTINUED

Figure 4.1

Figure 4.5

Figure 4.2

Figure 4.3

Figure 4.4

OPERATING INSTRUCTIONS

Open battery lid and install 
2 AAA batteries. Make 
sure the positive and 
negative poles are facing 
appropriate direction.

 1.

Keep distance at 2-3 
in. (5-8 cm) from 
thermometer probe to 
the middle of forehead, 
without coming into 
contact with the forehead 
directly.

 2.

Select measurement 
mode and click the 
measurement button to 
start taking temperature.

 3.

Reading will display on the 
screen after successful 
measurement.