This section includes a list of error messages and frequently
asked questions for problems you may encounter with your
SAMSUNG ARM BPM. If the products not operating as you think it
should, check here before arranging for servicing.
Approx.285g(Excluding the batteries)
Approx.159mm*50mm*27mm
4×AAA batteries,user manual
Battery powered mode: 6VDC 4×AAA batteries
About 22cm~42cm
Type BF applied part
WARNING: No modification of this equipment is allowed.
Power supply
Measurement mode
Oscillographic testing mode
Measurement range
Measurement perimeter
of the upper arm
Weight
External dimensions
Attachment
Mode of operation
Continuous operation
Degree of protection
Protection against
ingress of water
Accuracy
Normal working condition
Storage & transportation
condition
Software Version
Pressure:
5°C-40°C within ± 3mmHg (0.4kPa)
Pulse value: ±5%
Rated cuff pressure:
0mmHg~299mmHg (0kPa ~ 39.9kPa)
Measurement pressure:
SYS: 60mmHg~230mmHg (8.0kPa~30.7kPa)
DIA: 40mmHg~130mmHg (5.3kPa~17.3kPa)
Pulse value: (40-199)beat/minute
Device Classification
Battery Powered Mode:
Internally Powered ME Equipment
A temperature range of: +5°C to +40°C
A relative humidity range of 15% to 90%,
non-condensing, but not requiring a water vapour
partial pressure greater than 50 hPa
An atmospheric pressure range of:
700 hPa to 1060 hPa
Temperature: -20°C to +60°C
A relative humidity range of ≤ 93%, non-condensing,
at a water vapour pressure up to 50hPa
A01
PROBLEM SYMPTOM
CHECK THIS
REMEDY
No power
or Low
batteries
Error
message
Display
will not light up
Batteries are exhausted. Replace with new batteries
Insert the batteries
correctly
Batteries are inserted
incorrectly.
E 01 shows
E 02 shows
E 03 shows
Movement can affect the
measurement. Relax for a
moment and then
measure again.
E 04 shows
EEx,shows on
the display.
A calibration error
occurred.
Retake the measurement.
If the problem persists,
contact the retailer or our
customer service
department for further
assistance.Refer to the
warranty for contact
information and return
instructions.
or
shows
The monitor detected
motion, talking or the
pluse is too poor
while measuring.
Warning
message
“out” shows
Out of measurement
range
Relax for a moment.
Refasten the cuff and then
measure again. If the
problem persists, contact
your physician.
Refasten the cuff and then
measure again.
The measurement
process does not detect
the pulse signal.
Loosen the clothing on the
arm and then measure
again
The treatment of the
measurement failed.
Relax for a moment and
then measure again.
The cuff is not secure
or Inflatable abnormal.
IP22: The first number 2: Protected against solid
foreign objects of 12,5mm Ф and greater. The
second number: Protected against vertically falling
water drops when enclosure titled up to 15º.
Vertically falling drops shall have no harmful effects
when the enclosure is titled at any angle up to 15º on
either side of the vertical
19
18
TROUBLESHOOTING
SPECIFICATIONS