Technical Data Sheet
Technical Data’s from the ovens
TC-40 TC-100
Outer Dimension
Width mm
460
570
Height mm
501
631
Depth mm
527
660
Clearance distance from back wall
mm
250
250
Clearance distance from side wall
mm
250
250
Inner Dimension
Width mm
340
450
Height mm
370
500
Depth mm
330
460
Internal Volume
L
40
100
Shelf Standard/max
1/8
1/8
Max. Load per shelf
Kg
20
20
Weight (empty)
Kg
30
50
Temperature Range approx. 5 °C over Room temp to
°C
200
200
Temperature Variation
1)
at 50 °C
± °C
0.4
0.4
Temperature Variation
1)
at 100 °C
± °C
1.0
1.0
Temperature Variation
1)
at 150 °C
± °C
1.5
1.7
Temperature fluctuation
2)
at 150 °C
± °C
0.2
0.2
Heating up time
3)
to
70°C
Min 10 15
to 150 °C
Min
24
35
to 200 °C
Min
45
Recovering time after 30 sec door opening
at 100 °C
Min
4
7
Air changes (exhaust flap open)
at 100 °C
x/h
59
29
Power supply (±10%) 50/60 Hz
Volt
230
230
Nominal Wattage
Watt
1100
1100
Heat radiation
at 100 °C
Watt
145
230
at 150 °C
Watt
300
544
Equipment
Microprocessor -Temperature Controller LCD Display
Yes
Yes
Timer
Hours / Min.
0-999h 59m
0-999h 59m
Adjustable fan speed
%
60 - 100
60 – 100
Printer – Communication Interface RS 232
Yes
Yes
Adjustable Print Interval
Yes
Yes
Programming
Program / Step
50 / 15
50 / 15
Ramp function adjustable in steps of
°C
0.1
0.1
1) Measured with 3 temperature probes on horizontal level / divided in 1/3 of the chamber size
2) maximum temperature deviation in time for one temperature probe
3) to 98% of set temperature
All technical specification are specified for units with standard equipment at an ambient temperature of 25 °C (77 °F) and a voltage
fluctuation of ±10 %. The temperature data are determinated in accordance to following DIN 12880, part 2 respecting the recommended
wall clearances of 10 % of the height, width and depth of the inner chamber. All indications are average values, typical for units produced in
series. We reserve the right to alter technical specifications at all times without prior notice.
- 6 -
Copyright © SalvisLab / Edition 191107