5.3.
E-cal Module Error
Description
Warning message as shown below. It is caused by a mismatch between the frequency range of VNA and E-cal kit
Solution
This warning means the frequency range of E-cal kit can not cover the frequency range of the VNA. 2 solutions for this situation.
1) Please re-configure the VNA frequency range and ensure it will not be wider than the E-cal kit frequency range.
2) Replace the E-cal kit with a wider frequency range model
Then click [Refresh] button.
5.4.
Ecaldll Failure
Description
When start the software a notice box pops out.
Solution
It caused by a broken or missing Ecaldll.dll
Uninstall the software and re-install it.