background image

SAV2040X Electronic Calibration Kit Module

User Manual

Saluki Technology Inc.

Summary of Contents for SAV2040X

Page 1: ...SAV2040X Electronic Calibration Kit Module User Manual Saluki Technology Inc...

Page 2: ...address required LAN IP address required Calibration Ports 1 4 Compatibility The software is compatible with following VNA Keysight PNA N5242A N5244A N5245A N5247A N5230C Keysight ENA E5071C E5061B R...

Page 3: ...ual rests with Saluki Saluki Tech owns the copyright of this User Manual which should not be modified or tampered by any organization or individual or reproduced or transmitted for the purpose of maki...

Page 4: ...Software 10 2 1 Connection Window 10 2 2 Measurement Window 10 3 MacroEmbedded 12 4 Software Operation 19 4 1 User Interface 19 4 2 Operation 20 5 Troubleshooting 23 5 1 ConnectionFail 23 5 2 USB Con...

Page 5: ...ation 1 1 Software Installation Please kindly follow the steps below to install Electronic Calibration software on a VNA or a computer 1 Double click the software installation file A welcome page is s...

Page 6: ...6 909 602 109 Email sales salukitec com www salukitec com 3 Enter the user name and the company name then click Next 4 Select Installation directory then click Next 5 Create a shortcut folder then cli...

Page 7: ...886 909 602 109 Email sales salukitec com www salukitec com 6 Check all installation information in the Ready to install box If all information is OK click Next 7 Click Finish to complete the install...

Page 8: ...alukitec com 1 2 Ecal Kit USB DriverInstallation 1 Open the computer manage window a Click on the Windows button b Right click on Computer c Left click on Manage 2 Find Ecal USB Driver in Computer Man...

Page 9: ...rowse in the popped out window 5 Browse the Software installation directory mentioned in section 1 2 In it there are 2 folders XP folder contains the driver for XP system Win 7 folder contains the dri...

Page 10: ...ble please fill in the VNA GPIB address Click Connect button when the address is configured Connection states will be shown in Information area Information This area will show the connected VNA inform...

Page 11: ...e Ecal kit port automatically Ecal Module This area shows the Ecal kit information PromptInformation This area shows the warnings or errors If everything is OK this area shows None Measure DoneButton...

Page 12: ...luki E cal software provide Micro files for Keysight PNA and ENA series vector network analyzer These files are saved in the software installation folder This chapter will introduce how to embed the m...

Page 13: ...tec com Select a blank in macro title box on left then click Edit button 3 Enter macro name in Macro Title box in this case we name it SLK_ECAL For Macro Executable box click Browse and select Electro...

Page 14: ...14 Tel 886 909 602 109 Email sales salukitec com www salukitec com 4 Then the macro is embedded User may use the macro to do the calibration ENA series 1 Open E5061B Macro panel 2 Press Misc Setup...

Page 15: ...Email sales salukitec com www salukitec com 3 Click Network Setup enter network setting interface Then click Telnet Server make sure the state is ON open the network server function 4 Press Macro Set...

Page 16: ...tware Select the ECAL VBA file under the software installation directory the default installation path is C Program Files Electronic Calibration 3 3 2 E506B Click Open to open the project file Note If...

Page 17: ...909 602 109 Email sales salukitec com www salukitec com 7 In the Select Macro interface click Ecal SLK_Ecal to start e cal kit software 8 Then the macro is embedded User may use the macro to do the c...

Page 18: ...18 Tel 886 909 602 109 Email sales salukitec com www salukitec com...

Page 19: ...indows Connection window is used to build the connections Measurement window is used to configure the calibration details Connection window As shown in the figure below the interface for connecting th...

Page 20: ...on Reconnect After confirming that the ECal module is properly connected to the vector network reload the ECal module information Measure Perform calibration Done Complete the calibration process Quit...

Page 21: ...GPIB address set at the vector terminal and click the Connect button If the connection is successful go directly to the next step If the connection is not successful the display interface is as shown...

Page 22: ...vector network port 1 is connected to the calibration part B port B should be selected in the second list box of the first row You can check Auto Check Port to specify automatic port connection detect...

Page 23: ...ection fails Solution Check if the control computer and VNA are firmly connected Check the physical port of VNA computer is well contacted with the connector of the cable Check and confirm the IP GPIB...

Page 24: ...f E cal kit can not cover the frequency range of the VNA 2 solutions for this situation 1 Please re configure the VNA frequency range and ensure it will not be wider than the E cal kit frequency range...

Page 25: ...the connection cable Click Measure button and try again 5 6 Collection Lost between VNA and control computer Description Solution The software used SCPI to communicate with VNA VNA command error W R...