OPERATING INSTRUCTIONS
Operating the Unload Station
When the instrument is placed onto the unload station
and the power is on, it will automatically recognize its
presence. An audible alarm will notify the user that the
unload station is in place and the yellow LED on the [TRAY
UNLOAD] key will be illuminated.
After the installation procedures have been completed,
the instrument will already be powered on. The [POWER]
key LED should either be illuminated or flashing. If neither
is happening, check the power connection and the power
switch to ensure power is being supplied to the instrument.
If the [POWER] key LED is flashing, the instrument is in
the standby mode. Press the [POWER] key briefly to
return the instrument to the operational mode.
NOTE:
Every time the printer is powered on, the stack of
trays automatically moves up one tray, to ensure that the
new print job is started with an empty tray.
Removing and Replacing Trays
When all unload trays have been filled and indexed out
of the unload station, the [TRAY LOAD] yellow LED flashes,
the printer issues an audible alarm and the [ONLINE] key
extinguishes.
1. If all the trays have been indexed completely out of
the Unload Station, remove the trays and prepare to
load new trays.
If the trays have not been completely indexed
0
out, press the
[TRAY LOAD]
key (Figure 4-F) for
longer than one second.
Figure 4-F
2. Place one to ten trays on the unload station
positioned with the circle cutout over the sensor on
the right of the unload station (Figure 4-G). Trays are
positioned in the uppermost position.
Figure 4-G
NOTE:
Be certain that trays are stacked correctly before
attempting to raise or lower the unload station.
3. Press and hold the
[TRAY LOAD]
key for
approximately 2 seconds to index the trays down. The
instrument will recognize the number of trays and
load them appropriately.
4. Press the
[TRAY LOAD]
key briefly to index the trays
up one at a time.
Indexing Trays
Press the
[TRAY LOAD]
key briefly to index the trays
up one at a time.