(3)
Connect communication cable to the Prisma
®
in advance.
Now, the Prisma
®
is ready for linking with the Film
®
.
If an installation place does not allow easy access to the back side of the Prisma
®
, the
exhaust duct and the drain hose should be secured prior to linking with the Film
®
.
3-3.
Installation of Tissue-Tek
®
Film
®
To remove link cover located at the right side of the Film
®
:
(1)
Open loading door. Using Phillips screwdriver No. 2,
remove a screw that secure link cover to instrument
base.
(2)
Remove link cover from the instrument.
3-4.
Link
(1)
Place the Film
®
on the left-hand side of the Prisma
®
,
and then put the right edge of instrument base of the
Film
®
on installation bar attached to the Prisma
®
.