![Sakura 6134 Manual Download Page 16](http://html.mh-extra.com/html/sakura/6134/6134_manual_801574016.webp)
AF0-IF-013-00
16
7.
Handling of the linked Prisma
®
This section explains the information necessary to actually use the linked Prisma
®
. Please
also refer to each operation manual of Tissue-Tek
®
Prisma
®
and Tissue-Tek
®
Film
®
.
*
To send stained-slide baskets to the Film
®
during operation, see “7-1 Changing the
system setup.”
*
To know how the linked Prisma
®
is operating, see “7-2 Staining process when the link is
used.”
7-1.
Changing the system setup
The Prisma
®
allows selecting whether or not the stained-slide baskets are transferred to the
Film®. To transfer them to the Film®, it is necessary to change settings in the Utility
Menu.
(1)
Select “System Setup” in the Utility Menu.
(2)
Touch the button for “Coverslipper Link”
to change “Not in Use” to “In Use”.
Note:
When the link information is not installed on
the Prisma
®
by the link installation program,
the button for “Coverslipper Link” is not
displayed.
If you do not want to transfer baskets to the
Film
®
, although it is linked, select “Not in Use”.