background image

6.

Memory Backup:

This warmer is equipped with a

5-minute memory backup

. It provides backup power to the status indicator system for a

short period of time in the event that the power switch is shut off, or power cord is removed from the wall outlet. If power is interrupted, the

status indicators will remember the status of all the packages in the warmer at that time and reinstate those settings when power is restored

within 5 minutes.

7.

Premoistened cloth and hair care products manufactured by Sage Products Inc.:

Refer to individual product package for instructions,

ingredients, warnings and other information.

4.

Replacement parts and service:

Before performing any repairs or service on this warmer, contact Sage Products.

5.

Fuses:

This warmer is equipped with two externally accessible fuses. Turn off and unplug the warmer before attempting to remove fuses. If the

warmer stops running, check the fuses on the back side of the cabinet’s exterior. If a fuse is blown, replace it with another fuse of identical

rating. If the new replacement fuse blows, contact Sage Products for further instructions.

1.

Unplug the warmer before cleaning.

2.

Keep the inside of the warmer clean. If spatters or fluid adhere to walls, wipe with a damp cloth. Mild detergent may be used if the warmer

gets very dirty. The use of harsh detergent or abrasives is not recommended.

NEVER

clean the warmer with alcohol or solvent-based

cleansers. Alcohol or solvents on door(s) or status indicator display columns can weaken and cause cracking of these components.

3.

The outside warmer surfaces should be cleaned with a large, damp cloth. To prevent damage to the operating parts inside the warmer, water

should not be allowed to seep into the ventilation openings or any other openings on the cabinet exterior. Use cloths that have been

dampened with mild detergent solution only, but not so wet as to drip liquid or cause liquid to pool in warmer.

4.

If external electrical components become wet, including the display columns, power cord, power switch on front of cabinet and external

fuse access (if so equipped), clean with a soft, dry cloth and allow to air dry.

DO NOT

use harsh detergents or abrasives on any

external components.

5.

Cleaning inside the warmer outer panels and inside the warmer base should be performed only by qualified service personnel after taking

precautions to drain all stored voltage potential. Contact Sage Products for authorized service.

Adjusting the Temperature

The warmer is capable of warming products to a user-defined temperature. The user has the ability to adjust the temperature between 100°F and

125°F ± 5°F (35°C - 55°C).

1.

Press the MENU key below the display from the default temperature display screen.

2.

Using the UP/DOWN arrow keys located to the right of the display, highlight SET TEMPERATURE and press the SELECT key below

the display.

3.

When prompted, enter the following key sequence: UP ARROW, DOWN ARROW, UP ARROW.

4.

Using the UP/DOWN arrow keys, select the desired temperature for the contents of the warmer.

5.

After the desired temperature is selected, press the SAVE/EXIT key located below the display.

If you wish to discard the changes, press the CANCEL key located below the display.

6.

The warmer will then display the temperature set point on the main screen.

If the temperature of the warmer falls outside of the set point ± 5°F, the display will read the actual temperature of the warmer.

CARE OF YOUR WARMER

ACCESSING THE USER-DEFINED FEATURES

CAWR227

Summary of Contents for 7938

Page 1: ...warmer UL60601 1 Medical Electrical Equipment Part 1 Prior to the first edition of UL60601 1 general requirements for safety for the products covered by this standard were included in UL2601 1 This s...

Page 2: ...ug in water 13 Keep cord away from heated surfaces 14 The warmer is to be serviced only by qualified service personnel If there are no external or user serviceable fuses contact Sage Products for repa...

Page 3: ...e stocked If more than a few packages of a certain product type have the blinking TAKE FIRST arrow lit for an extended period of time it may be appropriate to cut down on the number of slots of that p...

Page 4: ...a grounded plug The plug must be plugged into an outlet that is properly installed and grounded WARNING Improper use of the grounding plug can result in a risk of electric shock Consult a qualified el...

Page 5: ...if no packages are present in warmer slots If packages are present the LED indicators for the filled slots will indicate NOT READY If a fault condition exists LED indicators will not light up if a pa...

Page 6: ...operating parts inside the warmer water should not be allowed to seep into the ventilation openings or any other openings on the cabinet exterior Use cloths that have been dampened with mild detergen...

Page 7: ...isplay 7 When the voice message is complete press the STOP key located below the display 8 The UP DOWN arrow keys can be used to review the message If satisfied with the message press the SAVE EXIT ke...

Page 8: ...A 5 V DC Y9897 Sage Products Inc 2009 IEC Standard IEC 60601 1 CAN CSA Standard CAN CSA C22 2 No 601 1 Characteristics Powder coated steel exterior aluminum interior surfaces and shelves Fuse Type 8 A...