Doc no: M18 Issue no: 07 Created: 24/08/2009
8
Contents of Packaging (as standard).
- AIRONE
™
Filtration Fume Cupboard complete
-
One envelope containing Prefilters (these may in some instances be fitted in the unit)
- One
bag
containing:-
-
Operational Safety Log Book.
- Operational
Manual.
-
Quality Control Certificate.
-
Mains Lead.
-
Allen Key.
Additional packages will include, if ordered, support stand, filters, ducting kits, etc.
Please note that for optimum performance it is advisable to site the AIRONE
™
Filtration Fume
Cupboards as directed in the Operational Safety Log Book (ref.: BS:7258).
Space for free air circulation must be provided around the AIRONE
™
Filtration Fume
Cupboard.
Place the AIRONE
™
Filtration Fume Cupboard on the selected bench or stand.
Fitting the Filters
It is advisable to have another person available to assist
The following procedure should be followed:
-
Ensure that power to the unit is switched off.
-
Ensure that the front sash is at its lowest position.
-
Unscrew the main front cover panel by removing the Allen screws to expose the main
filter housing aluminium cover plate, ensuring that the lighting plug is disconnected. If
possible have an assistant support the cover while this is done.
-
Loosen screws and remove cover plate by sliding it up and out of the keyhole slots.
- The filter housing units each comprise of a combined motor and filter plenum
chamber.
-
Unpack the Main Activated Carbon or HEPA filters and slide them gasket side
downward between the support guides for each main filter housing.