17
E1
Error Code
Cause
· Handpiece Overload
· Operation as condition
open the chuck
· If operation by the hand, press the motor button.
· If operation by the foot, release the foot controller.
· Operate after cooling for about 10 minutes
at the cool place.
· If the problem is not resolved, contact the manufacturer.
· Use after closing the chuck of handpiece.
· Defective handpiece assembly
· Defective bearing
· Defective motor
· Contact the manufacturer.
Remedy
E0
Error Code
Cause
Remedy
■
Error Code
- FORTE 100
- FORTE
100α,
FORTE 100
α
D
E1
· Defective connection of
handpiece
· Abnormal operation of motor
· Handpiece Overload
· Operation as condition
open the chuck
· Check the connect condition of power cord and
handpiece cord.
· If the problem is not resolved, contact the manufacturer.
· Re-turn on the power.
· If operation by the hand, press the motor button.
· If operation by the foot, release the foot controller.
· Operate after cooling for about 10 minutes
at the cool place.
· f the problem is not resolved, contact the manufacturer.
· Use after closing the chuck of handpiece.
·
Defective handpiece assembly
·
Defective bearing
·
Defective motor
· Contact the manufacturer.
Summary of Contents for FORTE 100
Page 1: ...INSTRUCTION MANUAL FORTE 200 200 200 D FORTE 400 400s FORTE 100 100 100 D FORTE 300 300 300 D...
Page 4: ...4 II III III OFF...
Page 12: ...12 1 2 3 2 35mm 1 0 2 3...
Page 16: ...16 E1 10 E0 FORTE 100 FORTE 100 FORTE 100 D E1 10...
Page 24: ...24 3 5 RPM X1 000 I O...
Page 26: ...26 B 1 2 3 1 2 A B 1 2 3 0 A...
Page 36: ...36 RPM X1 000 I O...
Page 38: ...38 1 2 3 2 35mm 1 0 2 3...
Page 48: ...48 I O RPM X1 000 1 0 2 3...
Page 54: ...54 1 h h 10 30 10 80 700 1 060 20 60 10 90 500 1 060 EU 2 CE EC REP SN...