41
OPERATION PROCESS
1. Put up the handpiece and turn on the control box.
(When the power is turned on, the beep sound is ring and the RPM is displayed on the indicator.)
2. Select the operation method by pressing the Hand
/
Foot Select Button. When the left light is on, hand is selected,
when the right light is on, foot is selected
.
3. Select the rotation direction by pressing the Rotation Direction Select Button.
When the left light is on, the forward direction is selected, when the right light is on, the reverse direction is selected.
■
How to use when select the hand
4. Push the motor button.
5. Control the desired speed by turning the speed control
dial.
6. Handpiece is stopped by re-pressing the motor button
.
■
How to use when select the foot
4. Control the desired speed by turning the speed
control dial.
5. Operate the handpiece by controlling the foot controller.
(Handpiece is worked within setting range.)
6. Foot controller is released, handpiece is stopped
.
- The first characters shown on the indicator when the power is turned on are software versions.
■
Maximum Speed Change Function
(
FORTE
300α,
FORTE 30
0αD
Only)
It is function to change the maximum speed from 40,000 rpm to 50,000 rpm.
After setting the speed control dial at ‘0’, Press the Hand
/
Foot Select Button and Motor Button. Maximum speed is enlarged to
50,000 rpm. Re-press the two buttons, Maximum speed is returned.
■
Overload Protective Function
If handpiece is overloaded, after about 5 seconds, beep sound is ring and the error cord ‘E1’ is displayed at the indicator and
handpiece is stopped. Hand
/
Foot Select Button is selected as the hand, this function can be disabled by pressing the motor button.
Re-press the motor button, handpiece can be used.
Hand
/
Foot Select Button is selected as the foot, this function can be disabled by releasing the foot controller.
Re-press the foot controller, handpiece can be used.
E1
Error Code
Cause
· Handpiece Overload
· Operation as condition
open the chuck
· Use after closing the chuck of handpiece.
· Defective handpiece assembly
· Defective bearing
· Defective motor
· Contact the manufacturer.
Remedy
E0
Error Code
Cause
Remedy
■
Error Code
- FORTE
300α, F
ORTE
300α
D
E1
· Defective connection
of handpiece
· Abnormal operation of motor
· Handpiece Overload
· Operation as condition
open the chuck
· Check the connect condition of power cord and handpiece cord.
· Re-turn on the power.
· If the problem is not resolved, contact the manufacture.
· Use after closing the chuck of handpiece.
· Defective handpiece assembly
· Defective bearing
· Defective motor
· Contact the manufacturer.
- FORTE 300
· If operation by the hand, press the motor button.
· If operation by the foot, release the foot controller.
· Operate after cooling for about 10 minutes at the cool place.
· If the problem is not resolved, contact the manufacturer.
· If operation by the hand, press the motor button.
· If operation by the foot, release the foot controller.
· Operate after cooling for about 10 minutes at the cool place.
· If the problem is not resolved, contact the manufacturer.
Summary of Contents for FORTE 100
Page 1: ...INSTRUCTION MANUAL FORTE 200 200 200 D FORTE 400 400s FORTE 100 100 100 D FORTE 300 300 300 D...
Page 4: ...4 II III III OFF...
Page 12: ...12 1 2 3 2 35mm 1 0 2 3...
Page 16: ...16 E1 10 E0 FORTE 100 FORTE 100 FORTE 100 D E1 10...
Page 24: ...24 3 5 RPM X1 000 I O...
Page 26: ...26 B 1 2 3 1 2 A B 1 2 3 0 A...
Page 36: ...36 RPM X1 000 I O...
Page 38: ...38 1 2 3 2 35mm 1 0 2 3...
Page 48: ...48 I O RPM X1 000 1 0 2 3...
Page 54: ...54 1 h h 10 30 10 80 700 1 060 20 60 10 90 500 1 060 EU 2 CE EC REP SN...