USER MANUAL - SIP POWER ION PUMP CONTROLLER
M.HIST.0109.23
Rev.1
Page
49
of
62
These bits are reserved for future use.
•
Bit 1 – Ethernet flag
If this bit is set, it means that Sip Power features a Ethernet connection.
•
Bit 0 – Display flag
If this bit is set, it means that Sip Power features a display on the front panel.
Bit 16:31 – Hardware revision number
This field is structured as follow:
•
Bit 15:8 – Hardware revision major number
This is the hardware revision major number.
•
Bit 7:0 – Hardware revision minor number
This is the hardware revision minor number.
Bit 32:47 – Software version
•
Bit 15:8 – Software version major number
This is the software version major number.
•
Bit 7:0 – Software version minor number
This is the software version minor number.
Bit 48:79 – Serial number
This field shows the device serial number.
Bit 80:111 – Output current
This field shows output current in nA units (sampled in the same time slice of VIN and VOUT).
Bit 112:127 – Output voltage
This field shows output voltage in V units (sampled in the same time slice of VIN and IOUT).
Bit 128:143 – Input voltage
This field shows current input voltage in dV units (sampled in the same time slice of VOUT and
IOUT).
Bit 144:159 – Reserved
This field is reserved for future use.
Bit 160:175 – Internal temperature
This field contains the device internal temperature reading provided in Kelvin degrees.
Bit 176:191 – Arcing events number
This field contains the number of arcing events since last start (or restart) command.
Bit 192:223 – Total working hours
This field contains the device total working hours (only the time spent supplying current).
Bit 224:255 – Time elapsed since last start
This field contains the number of seconds elapsed from the last start (or restart).
Bit 256:271 – Power supply status
This field is structured as follow:
•
Bit 15:13 – Reserved bits