USER MANUAL - SIP POWER ION PUMP CONTROLLER
M.HIST.0109.23
Rev.1
Page
45
of
62
LIFE_TIME_RESET_SECRET[55:48]
R/W
R R R R R R R R
Bit 47 46 45 44 43 42 41 40
LIFE_TIME_RESET_SECRET[47:40]
R/W
R R R R R R R R
Bit 39 38 37 36 35 34 33 32
LIFE_TIME_RESET_SECRET[39:32]
R/W
R R R R R R R R
Bit 31 30 29 28 27 26 25 24
LIFE_TIME_RESET_SECRET[31:24]
R/W
R R R R R R R R
Bit 23 22 21 20 19 18 17 16
LIFE_TIME_RESET_SECRET[23:16]
R/W
R R R R R R R R
Bit 15 14 13 12 11 10 9 8
LIFE_TIME_RESET_SECRET[15:8]
R/W
R R R R R R R R
Bit 7 6 5 4 3 2 1 0
LIFE_TIME_RESET_SECRET[7:0]
R/W
R R R R R R R R
Bit 63:0 – Life time reset secret
These registers contain the life time reset secret.
To reset the total working hours counter of the power supply you have to enable
critical step1
and
critical step2
bypass then write the life time reset secret in LIFE_TIME_RESET register.