DHD-3 Digital Hand Dynamometer – Operation
Reading Mode
1. Press
the
PWR
button to power on the DHD-3.
2. Press
the
READ
button to change into
Reading Mode
.
3. Press
the
HAND
button to select test results for either the
RIGHT
or
LEFT
hand. If
Auto Mode
test results are stored in memory, the
HAND
button
selects
Auto Mode
test results for display of the
LEFT
(fl ashing) hand and
then the
RIGHT
(f lashing) hand, after selecting manual test results for
display of the
LEFT
(not fl ashing) hand and then the
RIGHT
(not fl ashing)
hand.
4. Press
the
READ
button to display the
Test Number
and saved test result
for the selected hand. Record the test result, then press the
READ
button
again to display the next test result for the same hand in descending
Test
Number
order. Either the
DATE
and
TIME
for each individually saved test
result, or the
AVG
,
SD
and
CV
based on the total number of test results for
the selected hand are displayed sequentially by pressing the
SET
button.
5. Repeat steps 3 and 4 for the other hand.
6. Press
ZERO
once to erase an individually displayed
Test Number
and test
result. Press and hold
ZERO
to erase all manually and
Auto Mode
saved
Test Numbers and test results.