Rycom STX10 Manual Download Page 8

CONFIGURATION AND FUNCTION OF MENU

1. Setting the mode
a. Press Mode button, and the screen will display: Body...°F
b. Press again and the screen will display: Surface Temp...°F
c. Press again and the screen will display: Room...°F

Note: The thermometer default is set to BODY mode

2. Press the °C/°F button to transfer between Celsius and Fahrenheit
3. Press 

Mem

 (Memory) button, to display the last temperature, and allow for a 

   viewing of the last 32 measurements taken.

4. Recalibration of device via the F4 MENU

The Sanomedics Non-Contact Professional Thermometer is factory-set. Depending on
various skin types and thickness, there may be temperature differences. 

The F4 function will allow unit recalibration. To recalibrate, measure an individual’s 
temperature using a conventional thermometer and measure it concurrently using the 
Non-Contact Professional Thermometer. If there is a difference between the two 
readings, recalibrate the Non-Contact Professional Thermometer.

Instructions for recalibration:

Press the “MODE” and “°C/°F” buttons at the same time for 3 seconds. The screen will 
then display: ”F4”
Press “Mode” to increase 0.1°F, press “ °C/°F ” to reduce 0.1°F.
Press “MEM” button to save the setting.

In the cases of seasonal or environmental changes a verification and adjustment 
should be carried out.

5. Accessing temperatures stored in memory

To access stored scans (the last 32 scans are available), press “MEM” button, each 

time the “MEM” button is pressed the unit will scroll back to the previous measurement. 
To exit the Memory function do not press any buttons on the device for 5 seconds.

 

Important! 

In order to get an accurate BODY temperature, it is important that the 
thermometer is set to BODY mode.

Utilizing SURFACE or ROOM modes on the thermometer will result in an inaccurate 
core body temperature reading on a patient.

Likewise, if taking a surface temperature of an object or liquid, be sure to set the 
thermometer to SURFACE Mode. For an accurate room temperature reading, the 
thermometer must be set to ROOM mode. 

8

Summary of Contents for STX10

Page 1: ...Non Contact Thermometer Professional...

Page 2: ...ometer can also be used to measure the temperature of a baby bottle bath or room using the surface temperature function SAFETY PRECAUTIONS s OLLOW THE MAINTENANCE ADVICE STIPULATED IN INSTRUCTIONS s O...

Page 3: ...FFESIONAL THERMOMETER WORKS All objects solid liquid or gas emit energy by radiation The intensity of this energy depends on the temperature of the object The Non Contact Professional Thermometer is a...

Page 4: ...wered for three to four minutes which is a difficult task for young children to accomplish Axillary armpit temperature Although it may be easy to measure axillary temperature it has been proven that i...

Page 5: ...s 96 6 99 7 35 9 37 6 65 years 96 4 99 5 35 8 37 5 Practical considerations when taking a temperature s N ORDER TO ENSURE THAT PRECISE AND ACCURATE TEMPERATURE MEASUREMENTS ARE OBTAINED IT IS essentia...

Page 6: ...ehind the ear lobe although the primary and most optimal area is the above the right eye 6 Basic Instrument Layout Infrared Lens LCD Display Mode Scan Button Charging Dock Celsius and Fahrenheit Memor...

Page 7: ...seconds The screen will then display F4 Press Mode to increase 0 1 F press C F to reduce 0 1 F Press MEM button to save the setting In the cases of seasonal or environmental changes a verification an...

Page 8: ...gs recalibrate the Non Contact Professional Thermometer Instructions for recalibration Press the MODE and C F buttons at the same time for 3 seconds The screen will then display F4 Press Mode to incre...

Page 9: ...e in the upper right portion of the display window The red LED light will remain illuminated until the device is fully charged While charging and in the Charging Cradle the device will display the amb...

Page 10: ...rofessional use Its longevity is designed for 40 000 readings Accessories User manual in English Charging Cradle USB Power Cord AC Adapter Breakaway Safety Lanyard Maintenance of the Product The prote...

Page 11: ...her above 109 F 42 9 C in Body mode or above 244 4 F 118 C in Surface Temp mode THE SCREEN DISPLAYS THE MESSAGE LO When using the Non Contact Professional Thermometer the message LO can show on the sc...

Reviews: