HOW TO TAKE A TEMPERATURE
Aim at the FOREHEAD, over the right eye*, holding the unit approximately 2 inches (5 cm)
from the body. Activate the thermometer’s target finding feature, two red dots will appear
on the patient’s forehead. Move the thermometer gradually towards the patient until only
one red dot is visible – this is the optimal range for temperature measurement.
Press the Scan button to instantly display temperature.
* You can also take the temperature behind the ear lobe, although the primary and
most optimal area is the above the right eye.
6
Basic Instrument
Layout
Infrared Lens
LCD Display
Mode
Scan Button
Charging Dock
Celsius and Fahrenheit
Memory
2in
The reliability of the measurement cannot be
guaranteed if the temperature is measured
from parts of the body that are not recom-
mended (anywhere other than the forehead
above the right eye, or behind the earlobe).
Infrared Len
n
s
s
s
Infrare
over 60mm
24.6˚
Light beam B
Light beam A
INCORRECT
less than 60mm
24.6˚
Light beam A
Light beam B
INCORRECT
60mm
24.6˚
Light beam A and B focus into one point
CORRECT
X
X
Target Finder
LED A
Target Finder
LED B