RWC PLANAR 230 Manual Download Page 6

 

  mark on the body is between the two lugs on the cam. 

 

•  Refit the cap with the 40°C marking lining up with the index mark  

 

  on the body. 

 

•  Check that the cap moves between the maximum and minimum  

 

  marking.

Record the system conditions and valve settings so that they can be used 
for valve maintenance.

MAINTENANCE

If water conditions or installation conditions are severe, and the valve is slug-
gish in operation, it is possible that there is a build up of scale within the 
valve. the valve can then be stripped down and cleaned very easily.

1.  Isolate the hot and cold supplies and remove the main body of the   
valve from the installation.

2.  Remove the control knob and cam. 

3.  Remove the headwork using a spanner. If using an adjustable  

 

  spanner ensure it is correctly adjusted to avoid damage to the head 
  work. The headwork can now be removed. (Fig 5) 

4.  The thermostatic element and piston can be lifted out. This leaves    
  the return spring, which can then also be lifted out. Ensure you note  
  the orientation of all the components so that they can be    

 

  re-assembled in the correct order.

5.  With the spring removed the ‘O’ ring seal can be inspected,   

 

  cleaned and re-greased. If the ‘O’ ring is damaged it must be  

 

  replaced. 

6.  The components removed should be inspected and if any are  

 

  damaged they must be replaced.

7.  Line strainers should be removed from the inlet fittings and rinsed   
  in clean water. The pipework should then be flushed to remove any  
  contamination that has fallen back down into the pipe. 

8.  If it is thought that cross flow is taking place, the check valves  

 

  should be removed from the inlet fittings, rinsed in clean water and  
  then replaced. The check valves can be removed by using a pair of    
  long nosed pliers, pulling them evenly on one of the fins. Do not    
  attempt to lever the check valves out, as this will damage them.

After cleaning and re-assembly, the valve should be reset and tested.

Reviews: