RWC PLANAR 230 Manual Download Page 5

1.  Setting the preset temperature in the non-adjustable form.
 

•   Remove the cap and the cam. 

 

•   Refit the cap and, at approx. 50% flow rate, adjust the  

 

 

 

  temperature to 40°C. 

 

•  Remove the cap and refit the cam so that the index mark on the  

 

  body is between the two lugs on the cam. 

 

•  Refit the cap with the index mark on the body between the Max  

 

  and Min markings on the cap and check the temperature at the   

 

  maximum and minimum flow rates.  There should be no more    

 

  than a 3°C difference from the nominal temperature setting.

 

•  The cap should be locked in position.

2.  Setting the maximum temperature in the adjustable form with  
 

a maximum preset temperature. 

 

•  Remove the cap and the cam. 

 

•  Refit the cap and, at approx. 50% flow rate, adjust the  

 

 

 

  temperature to the required maximum temperature.  

 

 

 

•  Refit the cap with the figures on the cap indicating the maximum  

 

  temperature aligned with the index mark on the body. 

 

•  Mark on the body of the valve (use a felt tip or similar temporary  

 

  marker) the point between the Max and Min marks. 

 

•  Remove the cap and fit the cam so that the outer edge of the    

 

  clockwise lug line up with the temporary mark. 

 

•  Refit the cap with the maximum temperature figure aligned with  

 

  the index mark.  The cap should not be able to rotate  

 

 

 

  anti-clockwise. 

 

•  Check the maximum temperature is correct.

3.  Commissioning in an adjustable form with no preset maximum
 

temperature.

 

•  Remove the cap and the cam. 

 

•  Refit the cap and, at approx.. 50% flow rate, adjust the    

 

 

  temperature to 40°C. 

 

•  Remove the cap and refit the cam so that the index mark on the  

 

  body is between the two lugs on the cam. 

 

•  Refit the cap with the 40°C marking lining up with the index  

The inlet fittings can be stripped down  

 

 

 

 

and cleaned except for the non-return   

 

 

 

valve that, if removed, needs to be  

 

 

 

 

 

replaced by a new cartridge.

In-service tests and any necessary maintenance should be carried out 
on a regular basis, at least once a year. A record of the results should be 
retained for future checks.

 

Flow rate

INLET 
STRAINER

NON-RETURN
VALVE

1/4" TEST
PLUG

Reviews: