2. Description of the RTI Slit Camera
5
Setting Up the Slit Camera for the First Time
1.0B
RTI Slit Camera User’s Manual
Figure 2-3 The optional Tilting Device is ideal for use in mammography where the
reference position normally is not orthogonal to the breast support.
2.1.
Setting Up the Slit Camera for the First Time
Before you use your Slit Camera for the first time, mount the Slit Camera and the stand
to make sure that all mechanical parts fits together, and that no parts have been
damaged during transportation.
2.2.
Hardware and Specifications
All specifications can be changed without prior notice. RTI assumes no responsibility
for any errors or consequential damages that may result from the misuse or
misinterpretation of any information contained in these specifications.
2.2.1.
Slit Camera
Supporting standards
IEC 60336
Tube focal spot types
Radiographic, CT, Mammographic, Dental
kV range
18 - 150 kV
Enlargement factor E
0.8 - 5
E = n/m, where:
n = slit to detector distance
m = slit to focal spot distance
Measurement range
0.1 - 5 mm
2.2.2.
Dimensions
Slit holder size
97 x 34 x 6 mm
Slit dimensions
10 µm (as defined by IEC 60336)
Slit material
Tungsten (W, 99.9 %)