background image

6. Appendix A 

19 

A.1. Nominal Focal Spot Size - IEC 60336 

 

1.0B 

RTI Slit Camera User’s Manual 

6. 

Appendix A   

 
This appendix shows the tables for the nominal focal spot values for the different 
standards. 
 

6.1. 

A.1. Nominal Focal Spot Size - IEC 60336 

The publication IEC 60336 describes the different methods to measure the 
characteristics of the focal spot. In IEC 60336 the slit method is recommended to get 
accurate determinations of the focal spot dimensions. The other methods, pinhole and 
star pattern, are also described in IEC 60336 because they are often used by 
manufacturers and for clinical testing of X-ray tubes. 
 
The nominal focal spot value (f) is a dimensionless quantity defined by IEC and the 
permissible values of focal spot dimensions measured with the slit method is shown in 
table A-1. For instance, if the dimensions of the focal spot size measured with a slit is 
0.31 

 0.60 mm then, according to table A-1, the nominal focal spot size is 0.3.  

 
When the dimensions of the focal spot are measured with the slit method the length 
shall be multiplied with a factor of 0.7 if the nominal focal spot size is larger than 0.3. In 
IEC 60336 the motivation for this is that the distribution in radiant intensity over the 
length has relatively shallow shoulders, and the linear dimensions for length of the focal 
spot become larger compared to the width even though the MTFs for both width and 
length may be approximately equal. The reason for not using the multiplier 0.7 on tubes 
with nominal focal spot sizes below 0.3 is said to be that in that range the dimensions 
of the distribution of radiant intensity are more rectangular over both length and width. 
The reasons given above for multiplying the length of the focal spot with 0.7 for focal 
spot sizes larger than 0.3 implies special tube constructions which are out of date when 
today's X-ray tubes are concerned, especially modern mammography X-ray tubes.

  

 
To get accurate results of the focal spot size measurements with slit and film they shall 
be performed with an enlargement factor described in table A-2. This will not be a 
problem when the manufacturers do the measurements because they have the 
opportunity to dismantle parts of the X-ray unit (for instance the breast support). 
However, when the slit method is used in a clinical environment to measure the focal 
spot size on mammography X-ray tubes it may be impossible to get the magnification 
of the focal spot required. The nominal focal spot value is often less than 0.4 and the 
distance between the collimator and the breast support is not large enough to get an 
enlargement factor as large as three. 
 
When the slit method is used to estimate the focal spot size only (not MTF calculations) 
it is accepted, according to IEC 60336, to examine the slit images by eye trough a 
magnifying lens. The magnifying lens shall have a built-in graticule with divisions of 0.1 
mm and a magnification of 10.  
 
 
 
 

Summary of Contents for Slit Camera

Page 1: ...4707 001199 1 0B RTI Slit Cameara User s Manual 2015 02 11 RTI Slit Camera User s Manual English Version 1 0B...

Page 2: ...tandards and Compliances 6 2 3 1 Waste Electrical and Electronic Equipment WEEE 6 3 Theory of Focal Spot Reproduction 7 3 1 Theory 7 3 2 Why measure the focal spot 8 4 X Ray Tube Focal Spot Measuremen...

Page 3: ...the Slit Camera Congratulations to your purchase of the Slit Camera from RTI You have now in your hand a quality tool for focal spot size evaluation It has been carefully designed to meet the needs o...

Page 4: ...ft and without right the optional tilting device The Slit Camera can be rotated 90 degrees in the Slit Camera holder See figure 2 2 below The 90 degree rotation simplifies the measurement of length an...

Page 5: ...have been damaged during transportation 2 2 Hardware and Specifications All specifications can be changed without prior notice RTI assumes no responsibility for any errors or consequential damages tha...

Page 6: ...finance the cost of recovery from municipal collection points reuse and recycling of specified percentages per the WEEE requirements Instructions for disposal of WEEE by Users in the European Union Th...

Page 7: ...tics of the focal spot in both length and width directions Figure 3 1 Focal spot reproduction The grey part to the right shows an image taken with a film in the picture plane Note the lighter part in...

Page 8: ...y shape 3 2 Why measure the focal spot The main aspect of X ray imaging is to get images of objects A typical X ray focal spot is shown in Figure 3 2 As seen there are two peaks in the width direction...

Page 9: ...r of things that cause blurring of the X ray radiograms 1 Geometrical blurring a Focal spot size b Object to picture distance 2 Movement blurring 3 Material blurring Of these things the focal spot siz...

Page 10: ...er s Manual 1 0B Figure 3 4 X ray images for different sizes of the foci Figure 3 5 Reproduction of an object with different foci If the tube filament is worn out and about to break you can often see...

Page 11: ...e all objects between the focal spot and table e g compression device diaphragms cones etc which can be removed easily 2 Place the focal spot test stand on the imaging table with the slit direction pe...

Page 12: ...the slit holder measure the distance in mm between the front edge and the open part of the slit holder using a ruler X2 in the figure below 3 Calculate the magnification using the following formula 34...

Page 13: ...units of mm 2 One has to be able to get the intensity profile across the image of the focus slit In example below the image size has been calibrated by knowing the true size of the digital detector a...

Page 14: ...ge of the slit is horizontal the image has to be rotated 90 deg before measure with ImageJ This is done from the menu Image Transform Rotate 90 Degrees Right 4 4 1 5 Measure the distances X1 and X2 In...

Page 15: ...values on the left and right edge The cursor x value is show in the lower part of the window In this example the width is measured as FWHM Full Width Half Maximum and is 3 32 mm fm 3 32 mm 4 4 1 7 Cal...

Page 16: ...ffective focal spot FWFM Full Width Fifth Maximum For a bell shaped curve distance parallel to the abscissa axis between the points where the ordinate has one fifth of its maximum value Analogous to F...

Page 17: ...you should use standard magnification factors depending on the nominal focal spot size as shown in Table 5 1 below If you use the choice Full MTF you can set M as you like 1 10 Table 5 1 The standard...

Page 18: ...ENCE AXIS intersects with the ACTUAL FOCAL SPOT By convention the point of intersection forms the center of the EFFECTIVE FOCAL SPOT From the IEC 788 standard RMS Root Mean Square a proposed new metho...

Page 19: ...though the MTFs for both width and length may be approximately equal The reason for not using the multiplier 0 7 on tubes with nominal focal spot sizes below 0 3 is said to be that in that range the...

Page 20: ...45 0 45 0 65 0 4 0 40 0 60 0 60 0 85 0 5 0 50 0 75 0 70 1 10 0 6 0 60 0 90 0 90 1 30 0 7 0 70 1 10 1 00 1 50 0 8 0 80 1 20 1 10 1 60 0 9 0 90 1 30 1 30 1 80 1 0 1 00 1 40 1 40 2 00 1 1 1 10 1 50 1 60...

Page 21: ...length 8 16 reproduction 7 size 9 theory 9 width 8 18 Focal spot theory 7 FWFM 16 Geometrical blurring 9 IEC 336 7 D 1 Length See Focal spot length Line Spread Function 16 LSF 16 Magnification factor...

Reviews: