Measure
2021.5B
RTI Scatter Probe User's Manual
16
Measure
3.1 Measuring using Ocean
The following description assumes that Ocean is installed on a computer and started. For more information about Ocean, see
separate documentation.
Furthermore it assumes that measurement is performed in an X-ray environment using a diagnostic x-ray unit with open
collimation.
The first measurement
Normal Mode (Auto trig)
When the irradiation rate is expected to be higher than 5 µGy/h (or 10 µSv/h) during the entire measurement, Normal Mode can
be used. The Scatter Probe will then auto-trig when irradiation rate is above trig level.
1. Connect the RTI Scatter Probe to the computer with an USB cable to the USB port, see Product overview.
2. Start Ocean and wait for the Scatter Probe to connect.
3. Place the RTI Scatter Probe in a position about 1 m from the x-ray source with the detector surface facing the x-ray unit.
4. From the Backstage in Ocean, select Quick Check.
5. Select the Scatter Probe if the Find/select meter window appears.
6. Select type of measurement - In this example select
Air kerma rate