background image

 

 

 

 

11 

c)

 

When the hose is completely dry inspect the hose for signs of wear as described above. 

 

3.

 

Air Detection Sensor Replacement 

Inspection Frequency:

 The instrument will inform the user of possible air detection replacement requirements with the 

error code 

Replace Air Sensor. 

Air Detection Sensor Replacement Procedure:

 

a)

 

Refer to the error code corrective actions and try each suggested action. If the instrument still indicates the error 
code then the air detection sensor will require replacement. 

b)

 

Power down the instrument and remove the sample hose from the instrument sample inlet port. 

c)

 

Turn the instrument onto its face and remove the eight phillips head screws from the rear countersunk holes. 

d)

 

Carefully remove the instrument case front to expose the internal components. 

e)

 

Carefully lift the control circuit board and disconnect the air detection sensor harness from the circuit by pulling 
straight out of the circuit board connector. 

f)

 

Lift the air detection sensor housing straight up from the instrument off of the support pins and remove the 
sensor by unthreading. Discard the spent sensor. 

g)

 

Obtain a replacement air detection sensor, part number 1-16-1013-00-0, and install it into the housing by 
threading to hand-tight. 

h)

 

Replace the housing into the instrument by seating the housing through holes over the two case pins. 

i)

 

Route the sensor harness to the circuit board and plug it into the sensor harness header of the circuit by pressing 
straight in. The harness is keyed and will only fit securely in one direction. 

j)

 

Verify that the circuit board is properly positioned onto the mounting studs of the instrument case. 

k)

 

Secure the housing halves together using the eight phillips head screws. 

 

 

C. Spare Parts Listing 

Spare parts are available directly from the Neutronics Incorporated Sales or Service Department. To order spare parts 
contact Neutronics Incorporated by phone at (610) 524-8800 or by fax at (610) 524-8807. Have a shipping name, address 
and purchase order number ready as well as the model and serial number of the instrument and the part number of the spare 
part. 

 

Spare 

Part 

Description 

    Order 

Part 

Number 

User 

Instruction 

Manual 

    035-80979-00 

Sample 

Filter 

     026-80128-00 

Summary of Contents for ID900

Page 1: ...l rti rtitech com Web site www rtitech com OPERATION MANUAL ID900 Portable Refrigerant Identifier Manual No 035 80979 00 WARNING Liquid Samples will Damage Unit Do not operate if filter has red flecks...

Page 2: ...1 Power Connector 5 B Sample Hoses 6 1 R12 Sample Hose 6 2 R134a Sample Hose 6 C Power Harnesses 6 1 Vehicle Battery Adapter 6 2 Wall Power Adapter 6 D R134a Tank Adapter Fitting 6 E Storage Case 6 IV...

Page 3: ...e Input Warning The instrument requires connection of the sample hose to the low or vapor side port of the source vehicle or refrigerant cylinder Connection of the sample hose to the high or liquid po...

Page 4: ...a storage case that will house all components R12 and R134a sample hoses are provided as well as a R134a cylinder adapter fitting and a power cord for connection directly to the vehicle battery An op...

Page 5: ...as been analyzed by the instrument 7 Sample Inlet Port permits the connection of either a R12 or R134a sample hose to the instrument 8 Air Detection Sensor a chemical fuel cell that will detect the am...

Page 6: ...end This harness is utilized to connect power from the vehicle battery to operate the instrument and avoid the use of extension cords 2 Wall Power Adapter Optional is a 6 foot power cable that has an...

Page 7: ...ect the performance of the instrument Once the elevation has been entered it does not ever require entry again unless the instrument is moved to a new elevation The screen USAGE ELEVATION NOT SET is d...

Page 8: ...or Hydrocarbons have been detected in concentrations of 4 by weight or greater No weight concentrations of refrigerants or air are provided in this mode d NO REFRIGERANT CHK HOSE CONN Indicated whene...

Page 9: ...ensor failure due to temperature extremes Corrective Actions 1 Allow the instrument to stabilize at room temperature for at least 30 minutes before attempting use 2 To clear the error code power down...

Page 10: ...Procedure a Power down the instrument and disconnect the sample hose b Obtain a replacement filter c Remove the existing filter by pulling it straight up and out of the retaining clip and disconnecti...

Page 11: ...from the instrument off of the support pins and remove the sensor by unthreading Discard the spent sensor g Obtain a replacement air detection sensor part number 1 16 1013 00 0 and install it into the...

Page 12: ...erant R12 or R134a is detected in concentrations equal to or greater than 98 by weight Unknown Refrigerant Purity Setpoint In the presence of detected unknown refrigerant types R12 or R134a is conside...

Reviews: