background image

tetrahydrofuran, as well as unsaturated hydrocarbons, such as tetralin, diene, cumene and aldehydes, ketones 
and solutions of these substances.  

DANGER OF BURNING! The heating bath, tempering medium, evaporator piston and glass assembly can 

become hot  during operation and remain so for a  long time afterwards! Let the  components cool off before 
continuing work with the device.  

ATTENTION!  Avoid  delayed  boiling!  Never  heat  the  evaporating  flask  in  the  heating  bath  without 

switching  on  the  rotary  drive!  Sudden  foaming  or  exhaust  gases  indicate  that  flask  content  is  beginning  to 
decompose. Switch off heating immediately. Use the lifting mechanism to lift the evaporator piston out of the 
heating bath. Evacuate the danger zone and warn those in the surrounding area!  

•  When  the  device  is  switched  off  or  the  power  supply  disconnected,  the  internal  safety  lift  removes  the 

evaporator piston from the heating bath.  

CAUTION! Never operate the device when the evaporator piston is rotating and the lift is raised. Always 

lower  the  evaporator  piston  into  the  heating  bath  first  before  starting  the  rotation  drive.  Otherwise  hot 
tempering medium may be sprayed out!  

•  Set  the  speed  of  the  drive  so  no  tempering  medium  is  sprayed  out  as  a  result  of  the  evaporator  piston 

rotating in the heating bath. If necessary reduce the speed.  

• Do not touch rotating parts during operation. 

 

• Imbalance may result in uncontrolled re

sonance behaviour of the device or assembly. Glass apparatus may be 

damaged or destroyed. In the event of unbalance or unusual noises, switch off the appliance immediately or 
reduce the speed.  

• The appliance does not start up again automatically followin

g a cut in the power supply.  

• The device is only disconnected from the power supply network if the device power switch is off or the plug is 

pulled out.  

• The socket for the mains cord must be easily accessible. 

 

• The voltage stated on the type plate m

ust correspond to the mains voltage.  

• Socket must be earthed (protective ground contact). 

 

• Removable parts must be refitted to the appliance to prevent the infiltration of foreign objects, liquids etc. 

 

• Protect the appliance and accessories from bump

s and impacts.  

• The appliance may only be opened by experts. 

 

 
2. Proper Use 

 

The  instrument  is  designed  for  distilling  in  schools,  laboratories  or  factories.  Together  with  the  accessories 
recommended by manufacturer, the device is suitable for:  
- Quick and gentle distillation of liquids  
- Evaporation of solutions and suspensions  
- Crystallisation, synthesis or cleaning of fine chemicals  
- Drying of powder and granulate material  
- Recycling of solvents  
This device is not suitable for using in residential areas or other constraints mentioned in Chapter 1.  
 

3. Inspection 

 

3.1 Receiving Inspection 

 

Unpack  the  equipment  carefully  and  check  for  any  damages  which  may  have  arisen  during  transport.  Please 
contact manufacturer/supplier for technical support.  
 

Note: 

 

If there is any apparent damage to the system, please do not plug it into the power line.  
 
 
 
 
 
 

3.2 Listing of Items 

 

Summary of Contents for RS 100 PRO

Page 1: ...USER MANUAL EVAPORATOR RS 100 PRO ...

Page 2: ... Description of problem i e hardware or software Methods and procedures adopted to resolve the problems Your contact information Warranty This instrument is warranted to be free from defects in materials and workmanship under normal use and service for a period of 24 months from the date of invoice The warranty is extended only to the original purchaser It shall not apply to any product or parts w...

Page 3: ...of the drive There may be electrostatic discharges between the medium and the drive which could pose a direct danger The appliance is not suitable for manual operation Safe operation is only guaranteed with the accessories described in the Accessories chapter Refer to the operating instructions for the accessories e g vacuum pump Only use the device under an all side closed exhaust or a compa rabl...

Page 4: ...resonance behaviour of the device or assembly Glass apparatus may be damaged or destroyed In the event of unbalance or unusual noises switch off the appliance immediately or reduce the speed The appliance does not start up again automatically following a cut in the power supply The device is only disconnected from the power supply network if the device power switch is off or the plug is pulled out...

Page 5: ...The package includes the following items Items Qty Main unit 1 Heating bath 1 Power cable 1 Glassware vertical 1 User Manual 1 ...

Page 6: ...working state LED power Power ON LED power flash When the unit running LED power is lit Power switch Power ON or OFF Table 2 4 2 Display Display Descriptions Remote Display in case of remote control Probe Display when using external probe Mot Display when the rotating function is switched ON Temp Display when the heating function is switched ON Timer Display when timer function is switched ON Int ...

Page 7: ...Connect the device to the power supply using power cable b Connect the heating bath to the device c Figure 7 Remove the clamping device for the angle setting of the rotation drive on the right side of the lift by rotating the knurled screw d counter clockwise Set the drive at an angle of approx 30 Figure 8 Then secure the rotation drive from being accidentally turned by tightening the knurled scre...

Page 8: ... in connections Figure 9 Hold the tight fitting evaporator piston and counter clockwise e and then turn the plastic screw nut f until the evaporator piston neck Figure 9 Note Prior to commissioning hand tighten the plastic screw nut f left aligned Installing condenser seal Figure 10 and Figure 11 Figure 10 Figure 11 Installing condenser g Figure 12 and Figure 13 ...

Page 9: ...turn Tighten the cap nut hand screwed h Start the device at 200 rpm for 2 min Then tighten the cap nut firmly Figure 12 Figure 13 Installing the vertical glassware condenser locking device Figure 14 Figure 14 Connecting tubes Figure 15 Figure 15 5 2 Operation ...

Page 10: ...eck whether the fuse is broken or loose Fault in power on self test Switch OFF the unit then switch ON and reset the instruments to factory default setting If these faults are not resolved please contact manufacturer supplier 7 Maintenance and Cleaning Proper maintenance can keep instruments working properly and lengthen its lifetime Do not spray cleanser into the instrument when cleaning Unplug t...

Page 11: ...ment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense 9 Specifications Items...