background image

 

 

 

  

 
 

CO

2

 Incubator 

 

Model No: 050-600 

 
 

User Manual

    

 
 
 
 
 
 
 
 

 

 
 
 

Summary of Contents for Mini Galaxy E 050-600

Page 1: ...CO2 Incubator Model No 050 600 User Manual...

Page 2: ...5 Control Panel Description 8 6 Programming Temperature CO2 Auto Zeroing 9 7 Programming the Alarm System 10 8 The Alarm System 11 9 System Alarms 13 10 Cleaning and Disinfecting 14 11 Technical Spec...

Page 3: ...o control the level of CO2 giving excellent reliability and is unaffected by humidity The CO2 system has a semi automatic zero function which provides a simple process for the user to follow to mainta...

Page 4: ...midity tray cleaned with 70 Alcohol 30 distilled water solution every 10 14 days To remove replace the humidity tray turn the tray through 45 and remove for refilling To replace the humidity tray care...

Page 5: ...t are supplied fitted to the unit 4 Anti slip pads for the adjustable feet 1 Black sensor cover in situ If required further shelves can be purchased up to a maximum of 6 The humidity tray has a fixed...

Page 6: ...ully remove any packaging Fit 7 silicone rubber suction cup feet to the shelf racks in positions indicated on the picture below left with an Repeat for the opposite shelf rack To fit the shelf racks t...

Page 7: ...ray water should be changed and the humidity tray cleaned with 70 Alcohol 30 distilled water solution every 10 14 days To remove replace the humidity tray turn the tray through 45 and remove for refil...

Page 8: ...growth in the tray this can reduce contamination on the chamber walls one small teaspoonful copper sulphate recommended For IVF and other sensitive work we do not recommend the use of any biocide in...

Page 9: ...3 segment LED displays and 4 keys The functions and descriptions of these are described below PROGRAMMING Key PLUS Key MINUS Key ACCEPT Key The decimal points on the display for Temperature and CO2 f...

Page 10: ...ft causing an inaccuracy in the CO2 level at the programmed value It is recommended that the CO2 system is Auto Zeroed monthly to ensure that CO2 level is correct To carry out the Auto Zero press the...

Page 11: ...the display will show doo doo doo doo rAL rAL rAL rAL Press the Key and the Keys to adjust the time in the following steps 15 30 45 60 75 90 seconds then OFF Press to select the preferred value Press...

Page 12: ...pened the Alarm System is disabled and on closing the door the preset Alarm Arm timeout starts If chamber conditions recover within the Alarm Arm Time the Alarm System is rearmed At the timeout the Al...

Page 13: ...SE INNER DOOR 15 MINUTES ALARM DELAY TIMEOUT STARTED IS TEMPERATURE OR CO2 LEVEL MORE THAN 0 5 OUTSIDE PROGRAMMED POINTS CANCEL ALARM BY PRESSING ANY KEY ALARMS CANCELLED VISUAL AND AUDIBLE ALARMS CAN...

Page 14: ...ntrol valve The condition is shown on the display by the message Ct Ct Ct Ct riP riP riP riP When the chamber temperature has fallen to the programmed set point the system changes to mode 2 Mode 2 The...

Page 15: ...can be cleaned by rinsing in sterile water wiping down with the alcohol solution and rinsing with sterile water 10 9 Wipe down the inside of the chamber with alcohol water solution and leave to dry c...

Page 16: ...re to 50 C Control 0 1 C Stability 0 1 C Uniformity 0 2 C CO2 Control Solid state infra red CO2 sensor operating independent of humidity Programmable fully automatic zeroing facility Range 0 2 10 or 0...

Page 17: ...kg Electrical Supply Earthed Electrical Supply of Voltage 220 240v 50 60Hz Or 100 120v 50 60Hz Power 500 watts Energy to maintain 37 C 0 1kwh Storage Temperature 10 50 C NB Factory calibration of the...