1. INTRODUCTION 

This instrument is a digital thermometer for use with any K-
type thermocouple as a temperature sensor. 
Temperature indication follows the international temperature 
scale of 1990. (ITS-90) 

‰

 

Read the following information carefully before attempting to 
operate or service the meter. When servicing, use only 
specified replacement parts. 

Environmental conditions: 

c

 Altitude up to 2000 metres. 

d

 Relatively humidity 80% max. 

e

 Operating temperature: 0 to 40 °C 

U.S. Pat. No. 446,135

 

Safety symbols: 

 

Complies with EMC Directive 89/336/EEC 

 

2. SPECIFICATIONS 

2-1 Electrical specifications 

Measurement Range: -50 to 1300 

°

C

 ,

  -58 to 1999 

°

Resolution: 0.1°C, 1°C, 0.1°F, 1°F 

Basic Accuracy: (@23 ± 5

 Calibration) 

Function Range 

Accuracy  

± (% of r degrees)

°C

 

°C

 

°C

 

°F

 

°F

 

°F

 

-50

°C

 to 0

°C

 

0

°C

 to 1000

°C

 

1000

°C

 to 1300

°C

 

-58

°F

 to 0

°F

 

0

°F

 to 1832

°F

 

1832

°F

 to 1999

°F

 

0.5% ± 1

°C

 

0.3% ± 1

°C

 

0.5% ± 1

°C

 

0.5% ± 2

°F

 

0.3% ± 2

°F

 

0.5% ± 2

°F

 

 

1

Summary of Contents for 1319A

Page 1: ...RS DIGITAL THERMOMETER 1319A INSTRUCTION MANUAL MN AVG MX C F ...

Page 2: ...TABLE OF CONTENTS TITLE PAGE 1 INTRODUCTION 1 2 SPECIFICATIONS 1 3 NAME OF PARTS AND POSITIONS 3 4 OPERATION INSTRUCTIONS 5 5 OPEN THERMOCOUPLE INDICATION 7 6 RECALIBRATION PROCEDURE 7 7 MAINTENANCE 7 ...

Page 3: ...arts Environmental conditions c Altitude up to 2000 metres d Relatively humidity 80 max e Operating temperature 0 to 40 C U S Pat No 446 135 Safety symbols Complies with EMC Directive 89 336 EEC 2 SPECIFICATIONS 2 1 Electrical specifications Measurement Range 50 to 1300 C 58 to 1999 F Resolution 0 1 C 1 C 0 1 F 1 F Basic Accuracy 23 5 Calibration Function Range Accuracy of reading degrees C C C F ...

Page 4: ...eral specifications Power supply Qty 6 size AAA batteries Battery life approx 110 hours carbon zinc battery Low battery indication The symbol is displayed when the battery voltage drops below the operating voltage Sampling rate 2 5 times per second Weight 235 gms 8 29 oz Dimensions 5 91 L x 2 83 W x 1 38 H 150 x 72 x 35mm Operating temperature 0 to 40 C 32 to 104 F Operating humidity below 80 RH S...

Page 5: ...3 NAME OF PARTS AND POSITIONS MN AVG MX 6 9 5 1 3 2 10 T1 8 7 C F 4 MX AVG MN F C 3 ...

Page 6: ...nds to exit MX MN mode g ST Key Press S or T key to increase or decrease the READ mode memory location h MEM key Press MEM key each time stores one set of logged reading in memory i Ì key Press Ì key to enter the offset Relative mode zero the displayed reading as a reference value The offset mode annunciator OFFSET is displayed and the reference value displayed on the secondary display Press Ì key...

Page 7: ...ess C F key to the desired temperature scale f Perform measurements by contacting the object being measured with the probe sensor g Read the temperature on the display The display shows OL overload when the temperature being measured is outside the meter valid range 4 2 MAX MIN and AVG Function Operations c Press MX MN key to enter to MX MN mode to cycle through the maximum MAX minimum MIN or the ...

Page 8: ... down MEN key then press key to turn on the meter LCD will show ALL CLr mark press T key select or then press MEM key exit this mode If you select yes the all memory will be cleared 4 4 How to Disable Auto Power off Function The meter will enter to the sleep mode if no key pressed occurs for 30 minutes c Press key to turn off the meter d Press and hold down H HOLD key then press key to turn on the...

Page 9: ...rated once a year to ensure its continued accuracy Contact RS Components for further details of calibration service The address is given at the end of these instructions 7 MAINTENANCE 7 1 Cleaning Periodically wipe the case with a damp cloth and mild detergent Do not use abrasives or solvents Clean and dry as required 7 2 Battery Replacement When LCD display shows the batteries have insufficient p...

Reviews: