Datalogging Solar Power Meter / EN
7
08/15/16 Version No. 01
Calibration Factor Setting Mode : Push and hold this
key, when turning on the meter. The LCD displays
“CAL” and one flashing digit.
Push
, and keys to set up desired calibration
factor values. Push
↵
key to store and exit.
6. POWER / %
key :
Power and Transmission % Function Select : Push
this key to select power or transmission (%) function.
Power and Energy Units Select : In the integrated
solar energy mode, push this key to select power or
energy units.
Zero Adjustment : Push and hold this key for 3
seconds, when turning on the meter, the LCD will
show “CAL”. Cover the solar detector to zero the
reading then push
↵
key, the LCD will momentarily
display “MIN” to confirm the zero adjustment.
key : In READ mode, push this key to increase the
memory location.
In setting mode, push this key to increase the parameter.
7. MEMORY key :
Memory Function : Push this key momentarily to store a
measuring value.
Clear the stored Data : Push and hold this key, when
turning on the meter. The LCD shows “
CLr”.
Push
key to select “
” or “
”. to erase the
memory data. Push the
↵
key to confirm the selection.
key : In setting mode, push this key to move cursor
to the desired position.