Sensitivity Calibration
1. With the scale in regular weighing mode, press the
Tare
key to zero the
balance.
2. Press the
Cal
key.The balance will show “Scale”.
3. Press the
Tare
key. The balance will display one of the accepted
calibration weights.
Note:
The balance can only be calibrated in grams.
4. Press the
Mode
key to toggle between available calibration weights.
5. Press
Tare
. The balance will flash “0” briefly and then flash the
calibration weight.
6. Place the correct amount of weight (in grams) on the platform and press
the
Tare
key.
7. The calibration weight will continue to flash briefly and then become
stable.
8. Once the weight is stable, remove the weight from the balance. The
scale has now been calibrated and is in regular weigh mode.
9. Press the
Tare
key if the scale does not return to zero.
Linear Calibration
1.
2.
3.
4.
In regular weighing mode, press the
Tare
key to zero the balance.
Press the
Cal
key. The balance will show “Scale”.
Press the
Mode
key. The balance will show “Line”
Press
Tare
. The balance will flash “0” briefly and then flash the first
calibration weight.
5. Place the correct amount of weight (in grams) on the platform and press
the
Tare
key.
6. The balance will flash the next calibration weight.
7. Place the appropriate amount of weight on the weighing platter and
press
Tare
.
8. The calibration weight will continue to flash briefly and then become
stable.
9. Once the weight is stable, remove the weight from the balance. The
scale has now been calibrated and is in regular weigh mode.
10.Press the
Tare
key if the scale does not return to zero.
Summary of Contents for ES-1000H
Page 1: ......