Function Inspection
1. Press
DEFAULT
SETUP
to restore the oscilloscope to its default
setup.
2. Connect the ground alligator clip of the probe to the Ground
Terminal on the front panel.
3. Use the probe to connect the CH1 Input Terminal and theCompensation Sig
nal Output Terminal on the front panel.
4. Press “AUTO”.
5. Observe the waveform on the screen. In normal condition, the
display should be a square waveform as shown in the below:
6. Test the other channels in the same method. If the actual square
waveforms does not match that in the figure above, please
perform “Probe Compensation”.
Note
: To avoid electric shock when using the probe, please firstly
make sure that the insulated wire of the probe is in good
condition, and do not touch the metallic part of the probe
when it is connected to a high voltage.
Summary of Contents for 123-6438
Page 1: ...Quick Start RSDS1000CFL Series Digital Oscilloscope...
Page 2: ...2...
Page 7: ...7 152mm 118mm Appearance and Dimension 336 mm Figure 1 Front View Figure 2 Side View...
Page 32: ...32...
Page 33: ...33 Schnellstart RSDS1000CFL Serie Digitales Oszilloskop...
Page 34: ...34...
Page 57: ...57 D marrage rapide RSDS1000CFL Oscilloscope num rique...
Page 58: ...58...
Page 64: ...64 152mm 118mm Apparence et dimensions 336 mm Figure 1 Vue de face Figure 2 Vue de c t...
Page 89: ...89 Inicio r pido RSDS1000CFL Osciloscopio digital...
Page 90: ...90...
Page 96: ...96 152mm 118mm Aspecto y tama o 336mm Figura 1 Vista delantera Figura 2 Vista lateral...
Page 120: ...120...
Page 121: ...121 Avvio Rapido RSDS1000CFL Oscilloscopio Digitale...
Page 122: ...122...