Deep cleanse cycle
Yearly main incidentally
Deep cleanse medium
Chloride cleaning tablet
Exterior sanitation
Common lab practice
Maintenance
Preventive maintenance calls
per year
1, by qualified field service engineer
Consumables replacement
frequency
RPI: Every 3 months, WPU: Every 3 months
Consumables
RPI-Check / Progard / Q-gard / end point filters / peristaltic
pump tubing and pickup tubes
Electrical requirements
Mains voltage
120 or 230V (± 10%), 50-60 Hz (± 3%)
Fuse EU (230V)
2x 3.15A T (Slow blow) Hollyland 31-50T-032H or equivalent
according IEC 60127 (250V, temp range -55°C to + 125°C)
Fuse USA (120V)
2x 6.3 A T (Slow blow)Hollyland 31-50T-063H or equivalent
according IEC 60127 (250V, temp range -55°C to + 125°C)
Power consumption
600 VA
Protection class
IP 20
Overvoltage category
II
Pollution degree
2
External connections
RS232 Serial connection
RJ11
Ethernet connection
TCP/IP
External light (24V)(option)
24V, < 2A
USB
2.0
Environment
Performance characteristics and specifications
18