Page 9 of 16
4. Once the required unit has been selected, press the RESET button to move to the attached
totalizer.
5. The LCD will now display UNIT in the lower left-hand corner and TOTAL on the main display.
Press the MODE button to cycle through options available, as mentioned step 2.
6. Once the required unit has been selected, press the RESET button to move to the calibration
section.
Calibrating the Meter
1. The LCD will display CALIBRATE in the lower left-hand corner, as well as a number on the main
display. This represents the Test Volume to be dispensed through the meter during calibration.
1. Press the MODE button to cycle through options available.
• 2
• 4
• 8
• 20
• 100
• 250
2. Once the required Test Volume has been selected, press the MODE button for 3 seconds. The
LCD will display PURGE while CALIBRATE will also start to flash.
3. Purge the system of air by running fluid through the system. Press the MODE button to begin the
calibration process. The LCD will display RUN and the Test Volume (e.g., RUN 250).