RMS-Converter
E-M-RMS-Converter-V1_4.docx
Instruction Manual
© 2018 ROTRONIC AG Bassersdorf Switzerland
Page 96 of 126
Step 4
Data logging: the HL-NT has an internal memory, the HL-NT is battery powered. So the HL-NT can
log during a power interruption.
In case of a communication interruption to the RMS Server, the RMS Converter logs the data of the
HL-NT. After the interruption, RMS requests the data from the RMS Converter.
In case of a communication interruption between the RMS Converter and the HL-NT, the HL-NT
would log the data.
IMPORTANT: The HL-NT can log for a certain number of days. However, only the log values
of the past 7 days can be added into RMS via the RMS-CONVERTER, all other data will be
deleted!
Step 5
Sensor error: In case of one sensor error out of the two, RMS displays a sensor error for both
measurement values.
Step 6
Measurement alarm: Alarms linked to the measurement limits and programmed within the HC2 will
not be transmitted to RMS. However if the device has a display and the display is configured
correctly, the alarm will show on the display.
Step 7
Fix measurement values programmed within the Hygroclip are displayed as "Simulator" within RMS.
Chart (e. g. fix value for humidity)
(Temperature of the same Hygroclip is not marked as a Simulator)