2. A valve closes the “+” and “-“ connectors of the sensor internally and defines a new 0-point.
This procedure needs app. 10sec. The PF5 will not be able to send measurement values
out during this time.
Note: The PF4 has a 0-point compensation integrated within the device internal sensor. No
extra adjustment is necessary.
Note: A new 0-Point compensation overwrites the previous 0-Point compensation, but
doesn’t overwrites the current One point (≠0Pa) adjustment (see below).
Note: The real date and reference value cannot be shown due to missing compatibility of
these systems.
For relative humidity and temperature:
Humidity 2 points acquired
Humidity 2 points adjusted
Note: For humidity, multiple points can be acquired and adjusted. For temperature 1 or 2
points.
Step 4
Data logging: the PF4/5 has no internal memory, the PF4/5 is not battery powered. So the PF4/5
cannot log during a power interruption.
In case of a communication interruption to the RMS Server, the RMS Converter logs the data of the
PF4/5. After the interruption, RMS requests the data from the RMS Converter.
In case of a communication interruption between the RMS Converter and the PF4/5, the PF4/5
would not log the data.